a. Scale is off
b. Press and hold
TARE
, then press
ON/OFF
c. Scale displays
F.1
d. Press
TARE
four times
e. Scale displays
F.9
f. Press
MODE
to select between ¡Tr_on¡¨ or TroFF¡
-To enable
Auto Tare Function
, press
TARE
when
Tr_on
appears
-To disable
Auto
Tare
Function
, press
TARE
when
TroFF
appears
g. Press
TARE
for next
function setup or
ZERO
to restart
the scale
NOTE:
The Auto Tare Function will not be accessible if the
unit is purchased as legal for trade.
5. INSTRUCTION FOR USE
5.1 BEFORE WEIGHING
Make sure that:
a. Connect the load cell signal and power to the
indicator properly.
b. Place the scale on a level and firm surface.
c. The indicator is turned on.
d. The
ZERO INDICATOR
is on. If not, press
ZERO
to set
display to zero.
5.2 WEIGHING
a. Always place an object onto the scale gently.
Excessive force applied to platter may cause damage
to weight sensor.
b. The weight of the object is displayed on the indicator
automatically.
c. It is a good practice to remove all loads from scale
after weighing. It will prolong the lifetime of
weight sensor.
5.3 SELECT WEIGHT UNIT
Press
MODE
to shift between metric or avoirdupois weight
units. The weight unit used before power off will be
employed when the scale is turned on again.