V=00.00m/s 09Sec
Batt.Status=
100%
M OD IFY EQ UA TIO N
?
ENTER= YESC=EXIT
Opera onal Details For Indicator:
1.
Push the Power
On/Off
switch. The system performs a display test by glowing all display
segments of LCD & displays the name of the company.
2.
The display then performs a self-check process and displays the
Velocity .............. m/sec
with ba ery status and instrument is ready to take observa ons.
3.
Connect the both wires of Current meter sensor to the terminals provided on the top of
this Indicator.
4.
Lower the Current meter in to the water. A er 20 seconds (or entered Averaging me) it
directly displays the velocity of water in meters/sec.
5.
If the user wants to see/edit the calibra on equa on, then press the “FUNC” key.
6.
The display now displays the Calibra on equa on with the cursor blinking on the first
digit of the variable.
7.
User can change the digits by “UP” and “DOWN” keys. By pressing the “ENTER” key the
cursor will move to the next loca on.
8.
A er edi ng the values of variable and constant in the calibra on equa on press
“ENTER” key.
9.
Now cursor blinking on the first digit of the averaging period i.e. 20seconds. It
designates the me during which the velocity has to be averaged. It is adjustable from 1-99 sec
by “UP/DOWN/ENTER” keys as men oned above.
10.
Now instrument is ready to take observa ons as men oned above.
V=1.0000N+0.0000
AVG. TIME=10SECS
WATER CURRENT METER