Manual Data Printing (Reserved)

Manual: Under normal operation state, after unlocking, press and hold the down key 

for 3s to enter the data print interface;

Select dt1~dt7 by pressing the up and down keys, and press and hold the SET key 

for 3s to confirm the printing. Dt1~dt7 represents the current p

rinting time, which is 

pushed forward for 1~7 days. If there is no key operation for 1

0s, the interface will exit 

without saving.

Warehousing Mode

a. In the power-off operation state, after unlocking, press SET

 key +   key for more

than 3s at the same time, and the LCD display screen will display the parameter

code "PAS";

b. Press SET

 key to enter password setting;

c. Scroll the number to "22" with 

 key and   key;

d. Press the SET key to confirm the password. If the password is incorrect, the

warehousing mode setting program will exit;

e. If the password is correct, it will enter the warehousing mo

de and the power supply

of the temperature controller will be turned off;

f.

The above actions shall be completed during the buzzer alarm pe

riod (about 2

minutes) after power failure;

After entering the warehousing mode, the current consumption of the whole 

temperature controller is lower than 5uA.

Alarm Description

Code

Cause

Action

Err

Data access fault

N/A

E01

Biomedical freezer 

temperature sensor fault

The alarm symbol flashes; E01 is displayed; buzzer sounds 

immediately; compressor is in proportional operation; remote 

alarm is disconnected in mode C.

E02

Ambient temperature 

sensor fault

The biomedical freezer temperature sensor fails in A mode; 

there is no alarm in B mode and C mode;

AH

AH high temperature 

alarm

AH and temperature are displayed alternately; the alarm 

symbol flashes; the buzzer buzzes when A2 is exceeded; the 

remote alarm is disconnected in mode C.

AL

AL low temperature alarm

AL and temperature are displayed alternately; the alarm symbol 

flashes; the buzzer buzzes when A2 is exceeded; the remote 

alarm is disconnected in mode C.

EHi

Sensor temperature  

> 85°C

EHi and temperature are displayed alternately; the alarm 

symbol flashes.

ELo

< -50°C

ELo and temperature are displayed alternately; the alarm 

symbol flashes.

dor

Door open or not closed

The alarm symbol flashes; when 

A3 is exceeded, dor and 

temperature are displayed alternately and the buzzer buzzes; 

the remote alarm is disconnected in mode C.

N/A

Ambient temperature ≥ 

35°C

The alarm symbol flashes; the biomedical freezer temperature 

and ambient temperature are displayed alternately

.

N/A

Power-off

The alarm symbol is displayed; the buzzer sounds intermittently 

for 2 min.; the temperature is displayed intermittently for 20 h.

Maintenance

W

ARNING

 

In order to prevent electric shock or injury, the power supply must be cut off before 

any repair and maintenance of the equipment.

 

Make sure that drugs or suspended particles from the surrounding environment 

will not be inhaled during equipment maintenance. Otherwise it may cause harm 

to human health.

CAUTION

 

Do not spill water directly on the chamber body, so as not to cause the electrical 

components insulation performance decline and metal parts rust.

 

Do not use hot water and corrosive cleaners or organic solvents to clean the 

chamber body

.

 

Do not put heavy objects on top of the equipment to avoid deformation of the 

equipment under pressure.

Cleaning of Biomedical Freezer

The biomedical freezer should be cleaned once a month. Regular cleaning keeps

the biomedical freezer a brand-new appearance.

Wipe the dust off the biomedical freezer's shell, inner chamber

, and all

accessories with a dry cloth. If the biomedical freezer is dirty

, wipe it with a

cleaning cloth soaked with a neutral detergent to remove the dirt and wipe off the

residual detergent with a wet cloth. Then, wipe it with a dry cloth again.

Do not pour water onto the biomedical freezer shell or into the biomedical freezer

.

Otherwise, it may damage the electrical insulation, leading to a failure.

The compressor and other mechanical parts are completely sealed and do not

require lubrication.

Clean the frost or ice on the inner wall and the condenser filter screen once a

month.

Summary of Contents for VS-25L308

Page 1: ...EN 61326 1 2013 Product implementation standards EN 61010 1 2010 A1 2019 EN 61010 2 011 2017 VS 40L308 VS 25L106 VS 40L106 USER MANUAL MODEL VS 25L308 Freez...

Page 2: ...acoustic and optical alarm function allowing high and low temperature alarms power o alarm sensor fault alarm etc with alarm values adjustable on demand Design of safety door lock improves the safety...

Page 3: ...od ensure that children do not have access to the biomedical freezer The disposal of the biomedical freezer shall be accomplished by the appropriate personnel Doors shall be removed to prevent acciden...

Page 4: ...ical freezer Never store ammable explosive or volatile substances in the biomedical freezer or use ammable sprays near the biomedical freezer otherwise it may cause an explosion or re Do not leave the...

Page 5: ...t con rm that the temperature in the storage chamber has reached the set temperature and then put the items in batches Do not put in items for more than 1 3 of the volume of the chamber at a time to p...

Page 6: ...all be used to accelerate the defrosting process CAUTION Since the ambient temperature has a large impact on the equipment the equipment may not operate normally if the above environmental requirement...

Page 7: ...10 Product Components and Overview Schematic Diagram of Appearance Control panel Door Grip handle Control panel Door Exterior door handle Caster 11...

Page 8: ...SET key In the unlocking state press and hold the SET key for 3s to display the current control temperature set value St in the setting state press and hold the key for 3s to save and exit Up key to...

Page 9: ...con rm the password If the password is incorrect the parameter setting program will exit e Scroll the parameter names with key and key f Press SET key to display corresponding parameter values g Incre...

Page 10: ...d in mode C EHi Sensor temperature 85 C EHi and temperature are displayed alternately the alarm symbol ashes ELo 50 C ELo and temperature are displayed alternately the alarm symbol ashes dor Door open...

Page 11: ...rovided by the temperature recorder for use CAUTION Please install the temperature recorder by professional technicians or after sales personnel WARNING Please cut o the power supply before installing...

Page 12: ...220 240V 50Hz R290 118g Direct cooling Cyclopentane 308 195W 2 7kWh 24h 1 90A 102kg 700 690 1920mm Biomedical Freezer 25 10 40 20 Model No Climate type N N Class I Class I 220 240V 50Hz 220 240V 50Hz...

Page 13: ...ontroller USB board Door switch Ambient temperature sensor Biomedical freezer temperature sensor Condenser fan 22 Electrical Schematic Diagram Power plug Compressor Operating capacitor Alarm terminal...

Reviews: