Only professional technicians or maintenance personnel for after-sales 
support can disassemble and assemble the biomedical freezer. Otherwise, it 

Please use the special power supply indicated on the nameplate of the 

If the applied voltage is lower than 198V or higher than 242V, an automatic 
voltage stabilizer of more than 4000W suitable for motor load needs to be 
installed for assistance.

The power cord of this biomedical freezer is equipped with a three-prong 
(grounded) plug to mate with a standard three-prong (grounded) outlet of 
10A. Do not cut or remove the grounding pin of the power cord under any 
circumstances. Make sure that the power plug and the outlet are connected 

When the plug is removed from the power socket, it shall be held tightly, and 
the wire of the plug shall not be pulled. If the wire is pulled, electric shock or 

If the biomedical freezer is not operating properly, unplug the power plug. 
Continued operation under abnormal conditions may cause electric shock or 

Before any repair or maintenance of the biomedical freezer, always 
disconnect the biomedical freezer from the power supply to prevent electric 
shocks or personal injuries.

Make sure that the drugs or suspended particles inside and around the 
biomedical freezer will not be inhaled during repairs and maintenance, 
otherwise it may cause harm to human health.

When storing toxic, harmful, or radioactive substances, please use the 
biomedical freezer in a safe area. Improper use may cause harm to health or 
the environment.

Disconnect the power plug when the biomedical freezer is not in use for a 

the deterioration of the power cord.

If the biomedical freezer is to be stored unused in an unsupervised area for 
a long period, ensure that children do not have access to the biomedical 
freezer. The disposal of the biomedical freezer shall be accomplished by the 
appropriate personnel. Doors shall be removed to prevent accidents such as 

If CO /LN  standby system is used, a good ventilation environment must 

2

2

be provided at the installation and use site. Increased carbon dioxide 
concentrations in the air can asphyxiate and endanger life safety. In case 
of poor ventilation, other methods must be considered to ensure a safe 
environment.

Safety Precautions

Dear 

Vacc-Safe

 users:

We appreciate your trust. Thank you for using 

Vacc-Safe

 biomedical freezer. Please 

carefully read the following information to better understand the Operating 
Instructions and use this product, so as to prevent personal injury and damage to 
items.

Safety warning signs

Warning

Warning hands 

pinching!

Electrocution 

hazard!

Combustible 

material!

Safety Precautions

Document review is required in all 

cases marked with 

 to clarify the 

nature of potential hazards and the 
countermeasures that must be taken.

WARNING

Ignoring WARNING items may result in
serious casualties.

  Acts or operations that must 

be prohibited

CAUTION

Ignoring CAUTION items may result in 
casualties or damage to the biomedical 
freezer and related properties.

  Actions or operations that 

must be followed

WARNING

Please use a power socket with a grounding wire to prevent electric shock.
If the power socket is not grounded, the grounding wire must be installed by 
professional technicians.

proper care is taken to prevent overturning.

location, the biomedical freezer may fall down and cause injuries.

If the power cord needs to be extended, the cross-sectional area of the 

2

extension cord shall not be less than 2.5mm  and the length shall not be 

gas leakage, and open the doors and windows to ventilate and exhaust gas. 
Do not plug or unplug the power plug of the biomedical freezer, otherwise it 

Summary of Contents for VS-25L308

Page 1: ...EN 61326 1 2013 Product implementation standards EN 61010 1 2010 A1 2019 EN 61010 2 011 2017 VS 40L308 VS 25L106 VS 40L106 USER MANUAL MODEL VS 25L308 Freez...

Page 2: ...acoustic and optical alarm function allowing high and low temperature alarms power o alarm sensor fault alarm etc with alarm values adjustable on demand Design of safety door lock improves the safety...

Page 3: ...od ensure that children do not have access to the biomedical freezer The disposal of the biomedical freezer shall be accomplished by the appropriate personnel Doors shall be removed to prevent acciden...

Page 4: ...ical freezer Never store ammable explosive or volatile substances in the biomedical freezer or use ammable sprays near the biomedical freezer otherwise it may cause an explosion or re Do not leave the...

Page 5: ...t con rm that the temperature in the storage chamber has reached the set temperature and then put the items in batches Do not put in items for more than 1 3 of the volume of the chamber at a time to p...

Page 6: ...all be used to accelerate the defrosting process CAUTION Since the ambient temperature has a large impact on the equipment the equipment may not operate normally if the above environmental requirement...

Page 7: ...10 Product Components and Overview Schematic Diagram of Appearance Control panel Door Grip handle Control panel Door Exterior door handle Caster 11...

Page 8: ...SET key In the unlocking state press and hold the SET key for 3s to display the current control temperature set value St in the setting state press and hold the key for 3s to save and exit Up key to...

Page 9: ...con rm the password If the password is incorrect the parameter setting program will exit e Scroll the parameter names with key and key f Press SET key to display corresponding parameter values g Incre...

Page 10: ...d in mode C EHi Sensor temperature 85 C EHi and temperature are displayed alternately the alarm symbol ashes ELo 50 C ELo and temperature are displayed alternately the alarm symbol ashes dor Door open...

Page 11: ...rovided by the temperature recorder for use CAUTION Please install the temperature recorder by professional technicians or after sales personnel WARNING Please cut o the power supply before installing...

Page 12: ...220 240V 50Hz R290 118g Direct cooling Cyclopentane 308 195W 2 7kWh 24h 1 90A 102kg 700 690 1920mm Biomedical Freezer 25 10 40 20 Model No Climate type N N Class I Class I 220 240V 50Hz 220 240V 50Hz...

Page 13: ...ontroller USB board Door switch Ambient temperature sensor Biomedical freezer temperature sensor Condenser fan 22 Electrical Schematic Diagram Power plug Compressor Operating capacitor Alarm terminal...