Product Installation

Installation Environment

 

Ì

Ambient temperature: 10 °C~32 °C; ideal temperature: 18 °C~25 °C; air 
conditioning system shall be used if necessary.

 

Ì

Ambient humidity: below 80% RH. The humidity shall be less than 60% Rh if the 
maximum operating temperature is 32 °C.

 

Ì

Avoid large amounts of dust.

 

Ì

Avoid mechanical rocking or vibration.

 

Ì

Altitude of working position of biomedical freezer: lower than 2000m.

 

Ì

Input voltage: within 220V±10%.

 

Ì

The biomedical freezer is sensitive to ambient temperature. If installed in an 
environment other than the above, the biomedical freezer cannot operate 
normally. Please use it after improving the environment.

 

Ì

Do not install the biomedical freezer outdoors. Exposure to rain may cause an 
electric leakage or electric shock.

Installation Site

 

Ì

It shall not be installed in a narrow and closed space. The door body of the room 
where the biomedical freezer is placed shall not be smaller than or lower than 
the biomedical freezer, or at least the normal access of the equipment shall be 
ensured, so as to avoid maintenance di

翿

culties in the event of equipment failure 

and causing damage to the stored items due to failure of repairing the equipment 
in time.

 

Ì

The ground for installation must be solid and flat, non-combustible and able to 
withstand the weight of the equipment during operation.

 

Ì

Good ventilation, no direct sunlight.

 

Ì

Each refrigerator-freezer shall be provided with an independent power socket. 
Please ensure that the power socket shall carry a current of ≥ 10A and the plug 
shall be firmly connected to the socket. The power cord shall not be twisted or 
pressed.

 

Ì

Check the working voltage before use. A voltage stabilizer suitable for the 
equipment load shall be considered for use in areas with unstable voltage. The 
voltage stabilizer power shall be more than 4000W to ensure that the input 
voltage requirements in the installation environment are met.

 

Ì

The equipment shall be grounded securely. If the power cord socket is equipped 
with a grounding wire, check that the grounding is good before use. If the socket 
is not equipped with a grounding wire, be sure to have a professional engineer 
install the grounding wire.

WARNING

 

Ì

Do not connect the grounding wire to a gas pipe, water pipe, 
telephone line, or lightning rod when grounding the biomedical 
freezer, otherwise it may cause an electric shock.

 

Ì

The power plug must be accessible after installation to facilitate 
timely unplugging of the power cord in case of emergency. Do 
not block the vent of the biomedical freezer.

Precautions for Use

 

Ì

When the biomedical freezer is running, the left and right rear sides of the biomedical 
freezer may be heated, which is not a fault. In order to prevent condensation around 
the biomedical freezer housing, the condensate pipe is installed.

 

Ì

Before putting items into the equipment, confirm that the temperature in the 
storage chamber has reached the set temperature, and then put the items in 
batches. Do not put in items for more than 1/3 of the volume of the chamber at a 
time to prevent its temperature from rising too much.

 

Ì

The equipment temperature display is the temperature at the temperature sensor 
in the storage chamber. When the equipment is initially running, the displayed 
temperature is somewhat different from the actual temperature at the center of 
the equipment. However, as the equipment enters a stable state, the displayed 
temperature will gradually approach the actual temperature.

 

Ì

The equipment chamber is designed with a testing hole so that the test line 
inside the chamber can be led out for test. After the testing line is pulled out, 
it is necessary to plug the testing hole again with thermal insulation material, 
otherwise the temperature in the biomedical freezer may not reach the set 
temperature and cause condensation around the outside of the hole (applicable to 
biomedical freezers with testing holes).

 

Ì

Please use diluted neutral cleaner to clean the equipment. Do not use brushes, 
acid, gasoline, soap powder, polish or hot water to clean this equipment, otherwise 
it may damage the painted surface and plastic rubber parts. Special care must be 
taken not to use volatile solvents such as gasoline to wipe plastic and rubber parts.

 

Ì

When the biomedical freezer is not in use for a long period, cut off the power 
supply and turn off the battery switch.

 

Ì

If the biomedical freezer is not started again for a long time, a low battery level 
may occur. After the battery switch is placed at "ON", the battery can be charged, 
and the biomedical freezer can run continuously for about one week to charge the 
battery (applicable to models with batteries).

 

Ì

Please minimize the time of opening the door during each access to the items, 
so as not to cause great fluctuations in the temperature and humidity in the 
biomedical freezer.

 

Ì

After the door is opened, the temperature in the biomedical freezer will rise 
sharply in a short time, which is normal, and it will return to set temperature of 
normal operation after the door is closed.

 

Ì

After the biomedical freezer has been running for a period of time, a layer of frost 
is formed on the inner wall and evaporator of the biomedical freezer. If the frost 
layer is too thick, it will affect the thermal insulation effect of the biomedical freezer 
and increase the power consumption. Therefore, after the frost layer reaches 5 
mm, the attached shovel shall be used to defrost.

 

Ì

Before defrosting, the frozen items in the biomedical freezer shall be taken out 
and placed in an environment suitable for storage, so as not to damage the items 
due to temperature rise in the biomedical freezer.

 

Ì

There are many refrigerating coils in the biomedical freezer. The frost from 
the inner wall shall not be removed by sharp tools such as knife, ice chisel 
or screwdriver. The inner wall shall not be scratched during defrosting, or the 
biomedical freezer will fail.

Summary of Contents for VS-25L308

Page 1: ...EN 61326 1 2013 Product implementation standards EN 61010 1 2010 A1 2019 EN 61010 2 011 2017 VS 40L308 VS 25L106 VS 40L106 USER MANUAL MODEL VS 25L308 Freez...

Page 2: ...acoustic and optical alarm function allowing high and low temperature alarms power o alarm sensor fault alarm etc with alarm values adjustable on demand Design of safety door lock improves the safety...

Page 3: ...od ensure that children do not have access to the biomedical freezer The disposal of the biomedical freezer shall be accomplished by the appropriate personnel Doors shall be removed to prevent acciden...

Page 4: ...ical freezer Never store ammable explosive or volatile substances in the biomedical freezer or use ammable sprays near the biomedical freezer otherwise it may cause an explosion or re Do not leave the...

Page 5: ...t con rm that the temperature in the storage chamber has reached the set temperature and then put the items in batches Do not put in items for more than 1 3 of the volume of the chamber at a time to p...

Page 6: ...all be used to accelerate the defrosting process CAUTION Since the ambient temperature has a large impact on the equipment the equipment may not operate normally if the above environmental requirement...

Page 7: ...10 Product Components and Overview Schematic Diagram of Appearance Control panel Door Grip handle Control panel Door Exterior door handle Caster 11...

Page 8: ...SET key In the unlocking state press and hold the SET key for 3s to display the current control temperature set value St in the setting state press and hold the key for 3s to save and exit Up key to...

Page 9: ...con rm the password If the password is incorrect the parameter setting program will exit e Scroll the parameter names with key and key f Press SET key to display corresponding parameter values g Incre...

Page 10: ...d in mode C EHi Sensor temperature 85 C EHi and temperature are displayed alternately the alarm symbol ashes ELo 50 C ELo and temperature are displayed alternately the alarm symbol ashes dor Door open...

Page 11: ...rovided by the temperature recorder for use CAUTION Please install the temperature recorder by professional technicians or after sales personnel WARNING Please cut o the power supply before installing...

Page 12: ...220 240V 50Hz R290 118g Direct cooling Cyclopentane 308 195W 2 7kWh 24h 1 90A 102kg 700 690 1920mm Biomedical Freezer 25 10 40 20 Model No Climate type N N Class I Class I 220 240V 50Hz 220 240V 50Hz...

Page 13: ...ontroller USB board Door switch Ambient temperature sensor Biomedical freezer temperature sensor Condenser fan 22 Electrical Schematic Diagram Power plug Compressor Operating capacitor Alarm terminal...

Reviews: