background image

09

08

Operation after Power-off

The set value is memorized by the equipment. After recovery from a power failure, the 
equipment will resume operation with the setting parameters before the power failure.

WARNING

 

Ì

If the biomedical freezer is unplugged or the power to the 
biomedical freezer is interrupted, do not restart the biomedical 
freezer for at least 5 minutes to avoid damage to the compressor 
or system.

WARNING

 

Ì

When the biomedical freezer is not in use for a long period, 
unplug the power plug and turn off the battery switch (applicable 
to biomedical freezers with batteries) to prevent electric shock, 
electric leakage, or fire due to the deterioration of the power 
cord.

 

Ì

If the biomedical freezer is to be stored unused in an 
unsupervised area for a long period, ensure that children do not 
have access to the biomedical freezer and that doors cannot be 
closed completely.

 

y

The product shall be managed by a specially-assigned person to check 
the operation and make records every day (it is recommended to record 
the inspection once every 2~4 hours). In case of failure or shutdown, the 
temperature in the biomedical freezer will rise. If it cannot be repaired in 
a short time, the articles shall be transferred to other places that meet the 
temperature requirements for storage to avoid damage to the articles.

 

y

This product is a biomedical freezer. Please confirm in advance that the 
temperature required for the stored articles is consistent with the tem-
perature range of the biomedical freezer to avoid damage to the stored 
articles due to mismatch of the temperature range.

 

y

Due to the inertia of refrigeration, the temperature and humidity displayed 
for this product may differ from the actual temperature and humidity in-
side the chamber, which is a normal phenomenon.

 

y

Since all biomedical freezers are storage equipment, it is strictly prohib-
ited to put too many relatively hot items at one time, which will cause the 
compressor to not stop for a long time, the temperature will drop slowly 
and affect the service life of the compressor. The items must be put in 
batches to allow the temperature to decrease by steps until the required 
temperature is reached!

 

y

Do not damage the refrigeration circuit.

 

y

Do not use electrical appliances without production permits inside the 
biomedical freezer.

 

y

No mechanical tools or other means without the manufacturer's permis-
sion shall be used to accelerate the defrosting process.

CAUTION

 

Ì

Since the ambient temperature has a large impact on the 
equipment, the equipment may not operate normally if the above 
environmental requirements cannot be met. Please improve the 
environment before using the equipment. The equipment has an 
intermittent operation working system.

Preparations before Use

1. Remove all transport packaging materials and belts.
2. Counting the accompanying accessories. Check the items in the box against the

contents of the packing list. If there is any discrepancy, please contact with after-
sales service in time.

3. The placement conditions shall be such that clearances of at least 30cm are

reserved around the biomedical freezer for ventilation and heat dissipation.

4. Rotate the horizontal supporting feet clockwise with a wrench or manually to extend

downward so that they are supported on the ground to ensure that the biomedical
freezer does not move when used.

5. Take out the back bracket in the biomedical freezer, and fix the back bracket on the

back plate with screws.

6. Take out the handle, align the handle with the hole on the right side of the door, fix

the door handle to the door with an electric screwdriver, and then fasten the handle
trim cover to the upper and lower ends of the handle.

WARNING

 

Ì

Place the plastic packing bags in a place not accessible to 
children to avoid suffocation accidents.

First Power-up

Follow the steps below when using the equipment for the first time:
1. After the equipment is placed and leveled and cleaned, it shall be left for more than

24 hours before it is powered on to ensure normal operation.

2. Connect the power cord to a dedicated socket of suitable specification under no-

load condition.

3. After the power supply is turned on, turn on the power switch of the equipment first;

if the equipment is equipped with a battery, turn on the battery switch again.

4. Set the required temperature, do not put anything in the empty biomedical freezer, power

on and start up, check whether the operating temperature of the equipment reaches the
required temperature, observe the normal startup and shutdown of the equipment for
more than 24 hours, and demonstrate that the performance of the biomedical freezer is
normal.

5. Store items in batches, not exceeding 1/3 of the chamber volume at a time. Ensure

that the equipment is shut down and operates normally after being switched on and
off for more than 8 hours before placing the next batch of items.

6. Avoid opening the door during the cooling process as much as possible, otherwise

it will cause the temperature to rise.

Summary of Contents for VS-25L308

Page 1: ...EN 61326 1 2013 Product implementation standards EN 61010 1 2010 A1 2019 EN 61010 2 011 2017 VS 40L308 VS 25L106 VS 40L106 USER MANUAL MODEL VS 25L308 Freez...

Page 2: ...acoustic and optical alarm function allowing high and low temperature alarms power o alarm sensor fault alarm etc with alarm values adjustable on demand Design of safety door lock improves the safety...

Page 3: ...od ensure that children do not have access to the biomedical freezer The disposal of the biomedical freezer shall be accomplished by the appropriate personnel Doors shall be removed to prevent acciden...

Page 4: ...ical freezer Never store ammable explosive or volatile substances in the biomedical freezer or use ammable sprays near the biomedical freezer otherwise it may cause an explosion or re Do not leave the...

Page 5: ...t con rm that the temperature in the storage chamber has reached the set temperature and then put the items in batches Do not put in items for more than 1 3 of the volume of the chamber at a time to p...

Page 6: ...all be used to accelerate the defrosting process CAUTION Since the ambient temperature has a large impact on the equipment the equipment may not operate normally if the above environmental requirement...

Page 7: ...10 Product Components and Overview Schematic Diagram of Appearance Control panel Door Grip handle Control panel Door Exterior door handle Caster 11...

Page 8: ...SET key In the unlocking state press and hold the SET key for 3s to display the current control temperature set value St in the setting state press and hold the key for 3s to save and exit Up key to...

Page 9: ...con rm the password If the password is incorrect the parameter setting program will exit e Scroll the parameter names with key and key f Press SET key to display corresponding parameter values g Incre...

Page 10: ...d in mode C EHi Sensor temperature 85 C EHi and temperature are displayed alternately the alarm symbol ashes ELo 50 C ELo and temperature are displayed alternately the alarm symbol ashes dor Door open...

Page 11: ...rovided by the temperature recorder for use CAUTION Please install the temperature recorder by professional technicians or after sales personnel WARNING Please cut o the power supply before installing...

Page 12: ...220 240V 50Hz R290 118g Direct cooling Cyclopentane 308 195W 2 7kWh 24h 1 90A 102kg 700 690 1920mm Biomedical Freezer 25 10 40 20 Model No Climate type N N Class I Class I 220 240V 50Hz 220 240V 50Hz...

Page 13: ...ontroller USB board Door switch Ambient temperature sensor Biomedical freezer temperature sensor Condenser fan 22 Electrical Schematic Diagram Power plug Compressor Operating capacitor Alarm terminal...