Display Symbol

Meaning

Refrigeration symbol flashing: refrigeration on delay;

Refrigeration symbol normally on: the compressor is started and

 

refrigeration begins;

Power symbol normally on: the external power supply of the con-

troller is connected;

Lock symbol normally on: the controller is locked;

Alarm symbol flashing: the controller is in alarm state;

Defrost symbol flashing: defrosting dripping time;

Defrost symbol normally on: the controller is in defrosting state;

Controller Parameters and Operation

Under normal conditions, the controller displays and controls the biomedical freezer 

temperature measured. When the biomedical freezer temperature sensor fails, it 

displays "E01". When the biomedical freezer temperature is higher than 85°C, it 

displays "EHi". When the biomedical freezer temperature is lower than -50°C, it 

displays "ELo".

Control T

emperature Setting

In the unlocking state, press the SET key for more than 3s, and the display screen 

will display the current control temperature set value St; press the   key or   key to 

increase or decrease the value, and press the SET key for more than 3s to confirm 

the parameter value. If there is any parameter change, it will flash for 2s, and then 

store the modified parameter value and exit; otherwise, it will exit directly.

Setting of General Menu Parameters

a. In the normal operation state, after unlocking, press SET key +   key for more than

3s at the same time, and the LCD display screen will display the parameter code

"AH";

b. Scroll parameters (AH, AL, abc) with   key and   key;

c. Press the SET key to confirm the parameter, and the value of this parameter is

displayed;

d. Increase or decrease the value of the parameter with   key or   key;

e. Press SET key to temporarily store the modified parameter value and return to the

display parameter;

f. If the values of other parameters are modified, repeat steps b~e;

g. Press the SET key for more than 3s to store the modified parameter value and exit

the parameter setting program.

15

14

Management of Menu Parameter Setting

a. In the normal operation state, after unlocking, press SET key +   key for more than

3s at the same time, and the LCD display screen will display the parameter code

"PAS";

b. Press SET key to enter password setting;

c. Scroll the number to "-15" with   key and   key;

d. Press the SET key to confirm the password. If the password is incorrect, the

parameter setting program will exit;

e. Scroll the parameter names with   key and   key;

f. Press SET key to display corresponding parameter values;

g. Increase or decrease the value of the number with   key or   key;

h. Press SET key to temporarily store the modified parameter value and return to the

display parameter;

i. Repeat e~h if the values of other parameters are modified;

j. Press the SET key for more than 3s to store the modified parameter value and exit

the parameter setting program.

Note: If there is no key operation within 10s, the controller will automatically exit the 

parameter setting state (parameters are not saved) and return to the operation state. 

If there is no key operation for 3 minutes, the controller will be locked and the buzzer 

will buzz for 0.5s.

Press SET and FN for 3s at the same time to unlock or lock the locked controller, and 

the buzzer will buzz for 0.5 s. When the controller is locked, it cannot be operated. It 

can only be operated after unlocking;

T

ime Setting

The display shows "yer". At this time, the parameters "yer"/"mtH"/"day"/"HH"/'MM" 

can be selected through the up and down keys, representing year/month/day/hour/

minute respectively. After selecting the parameter to be set, press "set" to adjust 

the parameter. After adjustment, press and hold "set" for 3s to save and exit. During 

parameter adjustment, if there is no key operation within 10s, the setting will be exited 

directly, and all adjusted parameters will not be saved. The main control panel will 

buzz for 0.5s when the USB board is inserted, and the USB flash disk is inserted and 

pulled out.

If the recorder DR3180 is not connected or DR380 is abnormal, the display shows the 

character 'nDR' and the time cannot be set.

Export Data

Manual: Under normal operation state, after unlocking, press and hold the up key for 

3s to enter the data export interface;

If the recorder DR830 is not connected, the character "nDR" will be displayed; if the 

USB flash disk is not connected, the character "nUS" will be displayed;

Press up and down key to select M1~M12. Press and hold SET key for 3s to confirm 

the data export. M1~M12 represents the current time of data export, which is pushed 

forward by 1~12 months. If there are many data exported and the operation interface 

exits and the data is still being exported, "ULD" will be displayed at this time until the 

data export is completed. If there is no key operation for 3s, the interface will exit 

without saving.

Summary of Contents for VS-25L308

Page 1: ...EN 61326 1 2013 Product implementation standards EN 61010 1 2010 A1 2019 EN 61010 2 011 2017 VS 40L308 VS 25L106 VS 40L106 USER MANUAL MODEL VS 25L308 Freez...

Page 2: ...acoustic and optical alarm function allowing high and low temperature alarms power o alarm sensor fault alarm etc with alarm values adjustable on demand Design of safety door lock improves the safety...

Page 3: ...od ensure that children do not have access to the biomedical freezer The disposal of the biomedical freezer shall be accomplished by the appropriate personnel Doors shall be removed to prevent acciden...

Page 4: ...ical freezer Never store ammable explosive or volatile substances in the biomedical freezer or use ammable sprays near the biomedical freezer otherwise it may cause an explosion or re Do not leave the...

Page 5: ...t con rm that the temperature in the storage chamber has reached the set temperature and then put the items in batches Do not put in items for more than 1 3 of the volume of the chamber at a time to p...

Page 6: ...all be used to accelerate the defrosting process CAUTION Since the ambient temperature has a large impact on the equipment the equipment may not operate normally if the above environmental requirement...

Page 7: ...10 Product Components and Overview Schematic Diagram of Appearance Control panel Door Grip handle Control panel Door Exterior door handle Caster 11...

Page 8: ...SET key In the unlocking state press and hold the SET key for 3s to display the current control temperature set value St in the setting state press and hold the key for 3s to save and exit Up key to...

Page 9: ...con rm the password If the password is incorrect the parameter setting program will exit e Scroll the parameter names with key and key f Press SET key to display corresponding parameter values g Incre...

Page 10: ...d in mode C EHi Sensor temperature 85 C EHi and temperature are displayed alternately the alarm symbol ashes ELo 50 C ELo and temperature are displayed alternately the alarm symbol ashes dor Door open...

Page 11: ...rovided by the temperature recorder for use CAUTION Please install the temperature recorder by professional technicians or after sales personnel WARNING Please cut o the power supply before installing...

Page 12: ...220 240V 50Hz R290 118g Direct cooling Cyclopentane 308 195W 2 7kWh 24h 1 90A 102kg 700 690 1920mm Biomedical Freezer 25 10 40 20 Model No Climate type N N Class I Class I 220 240V 50Hz 220 240V 50Hz...

Page 13: ...ontroller USB board Door switch Ambient temperature sensor Biomedical freezer temperature sensor Condenser fan 22 Electrical Schematic Diagram Power plug Compressor Operating capacitor Alarm terminal...

Reviews: