5
and able to withstand the weight of this series of products in operation.
c) Placed in a place not exposed to direct sunlight and away from heat sources, at an ambient
temperature of not more than 32°C.
d) Require this series of products to be surrounded by a space of more than 30 cm to facilitate
ventilation and heat dissipation.
e) Must not be placed in an environment that is too cold to freeze;
f) Must not be placed in humid or splash-prone areas;
g) No electromagnetic interference in the surrounding area, which will affect the normal
operation of the control system and will directly damage the system in serious cases; Normal
operating conditions for this series of products:
a) Ambient temperature: 16℃~32℃;
b) Voltage range: 198V-264V;
c) Relative humidity: ≤ 80% RH;
d) No strong vibration and corrosive gases in the surrounding area;
e) Free from direct sunlight and other sources of heat and cold;
f) Altitudes below 2000 meters;
g) Overvoltage category II;
h) Pollution level (2)
The working system of this series of products: continuous working
WARNING: The packaging base at the bottom of this series of products must be
removed before use.
WARNING: When moving this series of products, take care that the bottom foot does
not touch the power cord.
WARNING: The bottom front wheel brake of this series of products need to be locked
after moving to the designated position.
WARNING: This series of products need to be left to stand for 12 hours before initial
power up.
WARNING: When this series of products are just connected to the power supply, do not
immediately put the items in to it. After running for a period of time, when the
temperature inside the freezer drops to the set temperature, then put the items in
batches.
Note: A single batch of items put into the freezer can not exceed 1/4 of the volume.
WARNING: Keep all ventilation openings in the enclosure or, in the structure for
building-in, clear of obstruction.
WARNING: Do not use mechanical devices or other means to accelerate the defrosting
process, other than those recommended by the manufacturer.
WARNING: Do not damage the refrigerant circuit.
For equipment that use flammable insulation blowing gases, the instructions shall
include information regarding disposal of the equipment.
Refrigerant and cyclopentane foaming material used for the appliance are flammable.
Therefore, when the appliance is scrapped, it shall be kept away from any fire source
and be recovered by a special recovering company with corresponding a qualification
other than being disposed of by combustion, so as to prevent damage to the