background image

10 

User settings are available after unlocked 
User menu parameter list 

Menu 

level 

Menu 

Menu description 

Setting range 

Default 

Unit 

User 

menu 

MAX 

Maximum temperature in the 

current time period 

--- 

--- 

 

MIN 

Minimum temperature in the 

current time period 

--- 

--- 

 

CLR 

Clear the record of the   

maximum and minimum   

temperature in the current 

time period. 

0: clear 

1: not clear 

SEt 

setting 

0.0~10.0 

 

Upper deviation on high 

temperature alarm 

0.0~10.0 

0: cancel the alarm 

 

Lower deviation on low 

temperature alarm 

0.0~10.0   

0: cancel the alarm 

 

Check the current date - year 

--- 

-- 

Check the current date - 

month 

--- 

-- 

Check the current date - day 

--- 

-- 

Check the current date - hour 

--- 

-- 

Check the current date - minute 

--- 

-- 

Pt 

Printing interval 

0~240 

20 

min 

tH1 

Ambient temperature alarm value 

20.0~50.0 

40.0 

 

P1 

Door heating mode selection 

1: automatic heating mode 1 
2: automatic heating mode 2 
3: automatic heating mode 3 

4: the door keeps open   

5: the door keeps closed 

P2 

Display mode selection 

1: average temperature 

2: upper temperature   

3: lower temperature 

PS1 

Password setting of user menu 

000~999 

005 

b1 

Hardware version 

--- 

1.0 

b2 

Software version 

--- 

1.4 

Summary of Contents for VS50

Page 1: ...keep it for future reference The design and speci cations are subject to change without prior notice for product improvement Consult with your dealer or manufacturer for details PHARMACY REFRIGERATOR...

Page 2: ...ries 6 3 PRODUCT OVERVIEW 7 4 TECHNICAL SPEC 7 5 OPERATION 8 5 1 Opera on and display panel 8 5 2 Key opera on 9 5 3 Noise Inside the Unit 13 5 4 ALARM 13 6 CLEANING 14 6 1 Cleaning the Interior and t...

Page 3: ...building in clear of obstruction WARNING Do not use mechanical devices or other means to accelerate the defrosting process other than those recommended by the manufacturer WARNING Do not damage the r...

Page 4: ...r more must be used and copper core conductors in the wall connected to the power socket must also have a cross sectional area of 4mm2 or more This series of products are strictly prohibited to put in...

Page 5: ...socket are plugged firmly and reliably otherwise it may lead to fires When unplugging from the power socket hold the power plug firmly and remove it Do not pull the wires of the power plug directly If...

Page 6: ...he insulation and cause leakage or electric shock Do not insert metal objects such as nails or wires into any openings and gaps of the Pharmacy Refrigerator or into any vent used for internal air circ...

Page 7: ...at both sides 15 cm at the rear and 20 cm at the top of the unit This will allow cold air to circulate around the Pharmacy Refrigerator and improve the efficiency of the cooling process The Pharmacy R...

Page 8: ...to the operation instructions below for details 4 Set parameter of cabinet temperature high low temperature alarm deviation etc according to actual need Default temperature setting is 5 C high temper...

Page 9: ...Model number VS50 External dimensions W D H mm 485 450 550 Effective volume 42L Door body Transparent glass door Foaming agent Cyclopentane Cooling mode Air cooling circulation Shell plate Inner Plat...

Page 10: ...on Button Locked off Button unlocked Power failure on Power off off Power on Printer disabled on Printer working off Printer not working Door heat disabled on Door heat working off Door heat not worki...

Page 11: ...locked if there is no operation for 60s or Up key and Down key are pressed for 3s at the same time b When the key is locked only ambient temperature can be checked and Mute key and Light switch key c...

Page 12: ...Lower deviation on low temperature alarm 0 0 10 0 0 cancel the alarm 3 n Check the current date year y Check the current date month r Check the current date day S Check the current date hour F Check t...

Page 13: ...o 1 Every time the door is opened and closed it will be heated for T9 minutes If the door is opened and closed again during the heating period the heating time will be reset Door heating mode P1 is se...

Page 14: ...eps once the digital tube displays the upload data code on in case of large amount of data and the USB flash disk generates the PDF files on data of current month and last month After the completion o...

Page 15: ...bient temperature alarm value tH1 sensor is normal and not shielded then the buzzer will alarm and the digital tube will flash the high ambient alarm code H2 alternately at the interval of 3 seconds e...

Page 16: ...fied technician for assistance 7 MAINTENANCE 7 1 Care When Handling Moving Your Unit Hold the unit around its sides or base when moving it Under no circumstances should it be lifted by holding the edg...

Page 17: ...t heat during operation specially in summer this is caused by the radiation of the condenser and it is a normal phenomenon Condensation condensation phenomenon will be detected on the exterior surface...

Page 18: ...16 REQUESTING ASSISTANCE OR SERVICE If you have any doubt or extra assistance please call Customer Service Vacc Safe Telephone 1300 459 140...

Reviews: