background image

13 

 

Battery is dead.   
 

Change battery now. 

 

Test strip may be damaged. 
 

Perform the test with new test strip. 

 

Test strip is used or test was 
not performed correctly. 
 

Perform the test with a new test strip and follow the test 
procedure correctly. 

No responses when 
strip is inserted into the 
meter 

Maybe: 
1. Battery is dead 
2. Wrong strip be inserted 
3. Meter is defective 

You have to: 
1. Replace battery 
2. Insert the test strip correctly 
3. Contact us for help 

No responses when 
sample is applied to the 
strip 

Maybe: 
Sample is not sufficient 
Meter is defective 

You have to: 
1. Repeat test with sufficient sample   
2. Perform Meter Check by inserting check strip. 

NOTE : If there is any error message displayed but not listed here, or your meter shut down 
for no reason, you can reset the meter by following two ways: 

1. Replace the battery.   

2. Open the battery door and use a needle to prink the 

“reset” hole (with a reset mark beside) 

the battery component.

Summary of Contents for GlucCell

Page 1: ...TM System consists of 1 the GlucCellTM Glucose Meter 2 GlucCellTM Glucose Test Strips and 3 the GlucCellTM code key and program key These products are intended to be used together to get accurate Cell Culture Glucose test results Do not use other commercial test strips with your GlucCell meter Important Information The GlucCellTM System is intended for in vitro analysis use with cell culture mediu...

Page 2: ...t BATTERY COMPARTMENT Hold one 3v Lithium battery The battery is not yet installed into meter when new purchasing Before using the meter please install the battery first TEST RESULT BATTERY SYMBOL Appear when battery is weak THERMOMETER SYMBOL Appears when ambient temperature is above or below the acceptable range needed for testing DROP SYMBOL Appear when the test strip is inserted and meter is r...

Page 3: ... used with samples containing no electrolyte Do not use test strips beyond the expiration date indicated on the strip vial label The discard date for test strips is 90 days after first opening the vial Record the discard date on the vial when you open a new vial of test strips GlucCellTM Glucose Test Strip could be damaged by heat and light Keep them sealed in the original vial Store the vial in a...

Page 4: ...ESSAGE PLEASE FOLLOW BELOW PROCEDURES 1 TURN OFF THE METER BY PRESSING M BUTTON FOR 3 SECONDS AND REMOVE THE CHECK STRIP FROM THE METER 2 INSERT CHECK STRIP TO REPEAT TEST IF THE SECOND RESULT PERSISTS CONSULT YOUR LOCAL DISTRIBUTOR FOR HELP 2 How to change measuring unit The meter has two measuring units to choose from mg dL and mmol L The default unit is mg dL if you do not change Step 1 Insert ...

Page 5: ...to turn on the meter The meter will enter to the setting mode automatically At the beep sound the year will flash for example for the year of 2009 number 09 will appear Press S or M button to obtain the desired year Wait the new setting flash 3 times to confirm and shifts to the next setting Repeat to set the month day hour and minute After minute is set the meter will display OK before turn off N...

Page 6: ... test strip vial FIG 2 If the two numbers match you may begin testing Otherwise continue to the next step Press S button until you hear the sound of buzzer and the code value flashes press S or M button to obtain the code number indicated on the new test strip vial FIG 3 Upon obtaining the right code number wait the new setting to flash 3 times to validate the change FIG 4 After screen shows the p...

Page 7: ...FIG 4 FIG 1 FIG 2 FIG 3 FIG 4 Carefully depress to form a 1 5uL droplet on the pipette tip Bring the droplet underneath the right or left aperture of the testing strip and touch gently to the strip allowing the entire droplet to be wicked into the strip FIG 5 Please allow the sample to be absorbed naturally to fill up the confirmation window Make sure that the sample has saturated the test confirm...

Page 8: ...Cell Culture Glucose result may be higher than 33 3 mmol L or 600 mg dL If LO is displayed your Cell Culture Glucose result may be lower than 1 1 mmol L or 20 mg dL FIG 9 When the measurement is higher than 500 mg dL please dilute the sample If the measurements is between 20 30mg dL the deviation may larger than 15 FIG 5 FIG 6 FIG 7 FIG 8 FIG 9 ...

Page 9: ...est result By pressing M button you will see your test record from the most recent to the oldest By pressing S button you will obtain the averages of the last 1 7 14 21 and 28 days NOTE ANYTIME IN MEMORY MODE YOU CAN PRESS M BUTTON UNTIL YOU HEAR THE BEEP TO EXIT AND TURN OFF THE METER To Delete Memory While in the memory mode select the test result you wish to delete and display it on the screen ...

Page 10: ...an 15 seconds 6 Measuring Range 20 600 mg dL 1 1 33 3 mmol L Alarm HI when over 600 mg dL LO when over 20 mg dL 7 Accurate Testing Range 30 500 mg dL 1 67 27 78 mmol L 8 Operating Temp Range 100 C 400 C 500 F 1040 F 9 Operating Relative Humidity 20 80 RH 10 Memory Capacity 180 test results with time and date 11 Power Supply One 3 volt Lithium Battery 12 Battery Life Approximately 1000 tests 13 Aut...

Page 11: ...y into the metallic hood of the battery holder with the plus side up Slide battery compartment door closed Check to see that your meter is working If it fails to turn on the battery may have been inserted incorrectly Remove the battery and reinsert it as illustrated NOTE EVERY TIME WHEN YOU REPLACE THE BATTERY THE METER WILL TURN ON AUTOMATICALLY AND ENTER TO THE TIME DATE SETTING PLEASE SET CORRE...

Page 12: ...ars contact your local distributor for help Cell Culture Glucose result is lower than 20 mg dL or 1 1 mmol L If this not confirmed the way you feel review proper testing procedure and perform a quality check with control solution Repeat test if LO still appears contact your local distributor for help Temperature is above or below the operating range of test strips The result you have obtained may ...

Page 13: ...u have to 1 Replace battery 2 Insert the test strip correctly 3 Contact us for help No responses when sample is applied to the strip Maybe Sample is not sufficient Meter is defective You have to 1 Repeat test with sufficient sample 2 Perform Meter Check by inserting check strip NOTE If there is any error message displayed but not listed here or your meter shut down for no reason you can reset the ...

Page 14: ...lucCell Meter and YSI Model 2300 Glucose Analyzer as reference The results are shown in the following table Slope 1 07 y intercept 5 9 mg dL 0 3 mmol L Correlation Coefficient R 0 99 Number of tests 120 Range tested 54 490 mg dL 3 0 27 2 mmol L SYMBOL Lot number Do not reuse Caution consult accompanying document Please consult instructions for use Use by Expiry date EU representative Temperature l...