Important information: Equipment marking (ATEX)
VACUUBRAND equipment bearing the ATEX mark
(see rating plate)
The classification according to ATEX is only valid for the inner part (wet-
ted part, pumped gas or vapor) of the equipment. The equipment is not
suitable for use in external, potentially explosive atmospheres (environ-
ment).
The overall category of the equipment depends on the connected com-
ponents. If the connected components do not comply with the classifi-
cation of the VACUUBRAND equipment, the specified category of the
VACUUBRAND equipment is no longer valid.
Vacuum pumps and vacuum gauges in category 3 are intended for con-
nection to equipment in which during normal operation explosive atmo-
spheres caused by gases, vapors or mists normally don’t occur; or, if they
do occur, are likely to do so only infrequently and for a short period only.
Equipment in this category ensures the requisite level of protection dur-
ing normal operation.
The use of gas ballast or the operation of venting valves is only permit-
ted if thereby explosive atmospheres normally don’t occur in the interior
of the equipment or, if they do occur, are likely to do so only infrequently
and for a short period.
The pumps are marked with ”X” (according to EN 13463-1), i.e., restric-
tions of the operation conditions:
• The equipment is designated for a low degree of mechanical stress
and has to be installed in a way so that it cannot be damaged from
outside.
Pumping units have to be installed so that they are protected against
shocks from the outside and against glass splinters in the event of
breakage (implosion).
Summary of Contents for MD 1
Page 27: ...page 27 of 72 MD 1 MD 1 C US pump with dual voltage motor 4 9 8 7 5 6 3 2 1 2 1 2 1 ...
Page 28: ...page 28 of 72 MD 1 VARIO SP Ensure sufficient venting of the pump 9 5 6 7 10 ...
Page 53: ...page 53 of 72 1 2 1 2 3 15 mm 4 6x size 2 2x 1 ...
Page 54: ...page 54 of 72 4 5 6 20696264 Clean ...
Page 55: ...page 55 of 72 Replacing the diaphragm 7 8 9 1 2 ...
Page 56: ...page 56 of 72 3 10 11 12 Pay attention to number of washers ...
Page 57: ...page 57 of 72 13 14 15 1 2 Pay attention to number of washers ...
Page 58: ...page 58 of 72 Replacing the valves and assembling the pump heads 16 17 18 4x ...
Page 60: ...page 60 of 72 22 23 16 15 24 1 2 Maintain all pump heads in the same way 15 16 mm size 2 2x ...