Abbreviations
FPM
: Fluoroelastomer
GND
: Ground
PE
: Polyethylene
PTFE
: Polytetrafluoroethylene
PWM
: Pulse width modulation
Position
Component
1
Power connection
2
ON/OFF switch
3
Voltage selection switch
4
Fuse holder
5
Pump rating plate
6
Inlet
7
Outlet
8
Fan
9
Handle (removable) / recessed grip
10
Control line
Pump parts
Gas inlet temperatures
* if pumping potentially explosive atmospheres: 50 °F to 104 °F (+10°C to +40°C)
Operating condition
Inlet pressure
Permitted range of gas
temperatures at inlet
Continuous operation
> 75 Torr (100 mbar)
(high gas load)
➨
50 °F to 104 °F
(+10°C to +40°C)
Continuous operation
< 75 Torr (100 mbar)
(low gas load)
➨
32 °F to 140 °F*
(0°C to +60°C*)
Short-time
(< 5 minutes)
< 75 Torr (100 mbar)
(low gas load)
➨
14 °F to 176 °F*
(-10°C to +80°C*)
Summary of Contents for MD 1
Page 27: ...page 27 of 72 MD 1 MD 1 C US pump with dual voltage motor 4 9 8 7 5 6 3 2 1 2 1 2 1 ...
Page 28: ...page 28 of 72 MD 1 VARIO SP Ensure sufficient venting of the pump 9 5 6 7 10 ...
Page 53: ...page 53 of 72 1 2 1 2 3 15 mm 4 6x size 2 2x 1 ...
Page 54: ...page 54 of 72 4 5 6 20696264 Clean ...
Page 55: ...page 55 of 72 Replacing the diaphragm 7 8 9 1 2 ...
Page 56: ...page 56 of 72 3 10 11 12 Pay attention to number of washers ...
Page 57: ...page 57 of 72 13 14 15 1 2 Pay attention to number of washers ...
Page 58: ...page 58 of 72 Replacing the valves and assembling the pump heads 16 17 18 4x ...
Page 60: ...page 60 of 72 22 23 16 15 24 1 2 Maintain all pump heads in the same way 15 16 mm size 2 2x ...