Replacing the device fuse (only MD 1 C/US)
➨
Unscrew the fuse hold-
er at the side of the
pump using a slotted
screw driver.
➨
Replace the defective
fuse by a fuse of the
same type (see „Tech-
Reassemble holder
with fuse to the pump.
➨
Switch off the pump.
➨
Disconnect the electrical power cord
before un-
screwing the fuse holder.
Identify and eliminate the cause of failure before switch-
ing on the pump again.
Summary of Contents for MD 1
Page 27: ...page 27 of 72 MD 1 MD 1 C US pump with dual voltage motor 4 9 8 7 5 6 3 2 1 2 1 2 1 ...
Page 28: ...page 28 of 72 MD 1 VARIO SP Ensure sufficient venting of the pump 9 5 6 7 10 ...
Page 53: ...page 53 of 72 1 2 1 2 3 15 mm 4 6x size 2 2x 1 ...
Page 54: ...page 54 of 72 4 5 6 20696264 Clean ...
Page 55: ...page 55 of 72 Replacing the diaphragm 7 8 9 1 2 ...
Page 56: ...page 56 of 72 3 10 11 12 Pay attention to number of washers ...
Page 57: ...page 57 of 72 13 14 15 1 2 Pay attention to number of washers ...
Page 58: ...page 58 of 72 Replacing the valves and assembling the pump heads 16 17 18 4x ...
Page 60: ...page 60 of 72 22 23 16 15 24 1 2 Maintain all pump heads in the same way 15 16 mm size 2 2x ...