Note
: In case of
supply voltage below 100V
, the lock
of the breaker may not latch and the pump might re-
start on its own after sufficient cooling. Take appropri-
ate precautions, if an automatic restart of the pump
may lead to a dangerous situation.
A warm up period (approximately 15 min.) is
required to ensure that the rated ultimate vac-
uum and pumping speed are attained.
O
➨ I
m
3
/h
Torr/mbar
✔
✔
~15min
Shutdown & storage
The pump can be switched off under vacuum.
Condensate?
O
➨ I
(5-10min)
atm
Short-term:
- Has the pump been exposed to condensate?
Allow the pump to continue to run at atmo-
spheric pressure for a few minutes.
- Has the pump been exposed to media which
may damage the pump materials or form
deposits
?
Check and clean pump heads if necessary.
Long-term:
- Take measures as described above regard-
ing short-term shutdown.
- Separate the pump from the application.
- Close inlet and outlet ports (e.g., with trans-
port caps).
- Store the pump under dry conditions.
NOTICE
NOTICE
air
Make sure the fan’s air supply is adequate.
Check fan regularly for dust/dirt. Clean fan
guard grill if necessary to avoid a reduction of
ventilation.
NOTICE
Summary of Contents for MD 1
Page 27: ...page 27 of 72 MD 1 MD 1 C US pump with dual voltage motor 4 9 8 7 5 6 3 2 1 2 1 2 1 ...
Page 28: ...page 28 of 72 MD 1 VARIO SP Ensure sufficient venting of the pump 9 5 6 7 10 ...
Page 53: ...page 53 of 72 1 2 1 2 3 15 mm 4 6x size 2 2x 1 ...
Page 54: ...page 54 of 72 4 5 6 20696264 Clean ...
Page 55: ...page 55 of 72 Replacing the diaphragm 7 8 9 1 2 ...
Page 56: ...page 56 of 72 3 10 11 12 Pay attention to number of washers ...
Page 57: ...page 57 of 72 13 14 15 1 2 Pay attention to number of washers ...
Page 58: ...page 58 of 72 Replacing the valves and assembling the pump heads 16 17 18 4x ...
Page 60: ...page 60 of 72 22 23 16 15 24 1 2 Maintain all pump heads in the same way 15 16 mm size 2 2x ...