Allow sufficient cooling of the pump before
starting maintenance.
+
Vent the pump and isolate it from the
vacuum system before you start mainte-
nance.
Ensure that maintenance is done only by suitably
trained and supervised technicians.
The valves and diaphragms as well as the motor capac-
itors are wear parts. If the rated ultimate vacuum is no
longer achieved or in case of increased noise level, the
pump interior, the diaphragms and the valves must be
cleaned and the diaphragms and valves must be checked
for cracks or other damage.
All bearings are encapsulated and are filled with long-life
lubricant. Under normal operating conditions, the drive
system is maintenance free.
In demanding circumstances, it may be efficient to
check and clean the pump heads on a regular basis. In
normal use, the lifetime of the diaphragms and valves
is typically 15,000 operating hours.
- Prevent internal condensation, transfer of liquids or
dust. The diaphragms and valves will be damaged if
liquid is pumped in significant amount.
- Carry out maintenance frequently if the pump is ex-
posed to corrosive media or in case of deposits.
- Regular maintenance will improve the lifetime of the
pump and also protect both users and the environment.
NOTICE
+
Service only one side of the pump at a time to avoid the mixing of
parts.
+
In case, support the pump appropriately.
atm
°C / °F
Service kit for MD 1 ..................................................................................
20696824
(diaphragms, valves, diaphragm key, silicone rubber tube for silencer)
Summary of Contents for MD 1
Page 27: ...page 27 of 72 MD 1 MD 1 C US pump with dual voltage motor 4 9 8 7 5 6 3 2 1 2 1 2 1 ...
Page 28: ...page 28 of 72 MD 1 VARIO SP Ensure sufficient venting of the pump 9 5 6 7 10 ...
Page 53: ...page 53 of 72 1 2 1 2 3 15 mm 4 6x size 2 2x 1 ...
Page 54: ...page 54 of 72 4 5 6 20696264 Clean ...
Page 55: ...page 55 of 72 Replacing the diaphragm 7 8 9 1 2 ...
Page 56: ...page 56 of 72 3 10 11 12 Pay attention to number of washers ...
Page 57: ...page 57 of 72 13 14 15 1 2 Pay attention to number of washers ...
Page 58: ...page 58 of 72 Replacing the valves and assembling the pump heads 16 17 18 4x ...
Page 60: ...page 60 of 72 22 23 16 15 24 1 2 Maintain all pump heads in the same way 15 16 mm size 2 2x ...