page 26 of 76
Technical data
General technical data valid for all pumps
ATEX approval if the ATEX marking
is shown on the rating plate
Inner part (pumped gases)
II 3/- G Ex h IIC T3 Gc X
Internal Atm. only
Tech.File: VAC-EX02
Maximum permissible inlet pressure
(absolute)
psi
(bar)
16
(1.1)
Maximum permissible outlet pressure
(absolute)
psi
(bar)
16
(1.1)
Maximum pressure difference between
inlet and outlet
psi
(bar)
16
(1.1)
Maximum permissible pressure (abso-
lute) at gas ballast valve
psi
(bar)
17.5
(1.2)
Permissible ambient temperature
storage / operation
°F
(°C)
14 to 140 / 50 to 104
(-10 to +60 / +10 to +40)
Permissible relative atmospheric mois-
ture during operation (no condensation)
%
30 to 85
Maximum permissible installation altitude
above mean sea level
ft
(m)
6500
(2000)
Rated motor power
hp
(kW)
0.71
(0.530)
No-load speed 50/60 Hz
rpm
30 - 2400
Maximum permissible range of supply
voltage ( ±10% )
Attention: Observe specifications of rat
-
ing plate!
100-120 V~ 50/60 Hz,
200-230 V~ 50/60 Hz
Device fuse
100-120V: slow blow fuse 8A
200-230V: slow blow fuse 4A
Motor protection
thermal cutout, manual reset
Overvoltage category
II
Degree of protection IEC 60529
IP 20
Degree of protection UL 50E
type 1
Pollution degree
2