Assembly of a vacuum system
+
Assemble vacuum connection lines between vacuum pump, CVC
3000 and vacuum application.
+
Assemble the vacuum connection line between the controller and the
vacuum application, if the controller is not part of a pumping unit.
+
Assemble electrical connections.
+
Connect coolant if necessary.
Vacuum for filtration and suction
+
Select ”
Pump down
” function.
+
If necessary, set value for
”Speed”
(high or low pumping speed). Re-
duce the speed in case of fast sedimentation (formation of a filter cake).
+
Set ”
Minimum
” to a value which provides adequate suction but does
not lead to evaporation of the solvent.
+
Start process by pressing ”START/STOP” key.
alternatively:
+
Select ”
Vac control”
function.
+
Set ”
Set vacuum
” to a value which does not lead to evaporation of the
solvent.
+
If necessary, set value for
”Speed”
(high or low pumping speed). Re-
duce the speed in case of fast sedimentation (formation of a filter cake).
+
Set
”Maximum”
pressure setting so that pump will switch off at the end
of the filtration process, or in the event that a filter cake cracks, leading
to a sudden pressure increase.
+
Start process by pressing ”START/STOP” key.
+
Tip for filtration:
Adjust preset pressure to a value well above the
boiling pressure of the solvent (e.g., for water >>15 Torr (20 mbar)).
Set the maximum pressure to e.g., 375 Torr (500 mbar). Once the fil
-
tration has finished, the pressure increases and the pump is switched
off automatically.
Application examples
Summary of Contents for MV 10C NT VARIO o.C.
Page 34: ...page 34 of 152 1 8 2 9 18 PC 3012 NT VARIO EK Peltronic 12 11 16 8 17 3 4...
Page 104: ...page 104 of 152 2 3 size 1 2x 4...
Page 105: ...page 105 of 152 6 5 7 4 4x...
Page 106: ...page 106 of 152 9 8 1 2 2 5 mm 2 5 mm 10 ME 16C NT VARIO MD 12C NT VARIO MV 10C NT VARIO...
Page 108: ...page 108 of 152 12 Emission condenser Peltronic 4 4x 11A 13 4 2x...
Page 109: ...page 109 of 152 15 14 4 4x Don t loosen this screw 15A 16...
Page 110: ...page 110 of 152 16 17 5 2x 15A 1 2 3 1 4 1 4 3 2x 5...
Page 111: ...page 111 of 152 19 20 18 5 2x...
Page 113: ...page 113 of 152 21 2 5 mm Replacing the diaphragms 22 23 696867 696839 Clean 5 8x...
Page 114: ...page 114 of 152 24 25 26 1 2...
Page 115: ...page 115 of 152 27 3 28 29 Pay attention to number of washers...
Page 118: ...page 118 of 152 36 37 38 Clean...
Page 119: ...page 119 of 152 39 40 41 IN IN EX EX 696867 696839 4x...
Page 120: ...page 120 of 152 42 43 44 IN EX T20 4x 2 2 ft lbf 3 Nm...
Page 121: ...page 121 of 152 45 46 47 Maintain all pump heads in the same way...
Page 125: ...page 125 of 152 2 size 2 2x 3 4A Version A valve 24 4 mm...
Page 126: ...page 126 of 152 5A 696839 24 4 mm 26 mm 4B 5B 696839 37 mm 40 mm Version B valve 37 mm...
Page 127: ...page 127 of 152 6 7 size 2 2x 7A ME 16C NT VARIO MD 12C NT VARIO MV 10C NT VARIO size 2 2x...
Page 129: ...page 129 of 152 49 Assembling the pump housing 50 51 5 2x 8 9 ft lbf 12 Nm...
Page 130: ...page 130 of 152 52 53 54 5 2x 8 9 ft lbf 12 Nm...
Page 131: ...page 131 of 152 55 55A 1 2 3 3 2x 55A 56...
Page 132: ...page 132 of 152 56 4 4x 3 7 ft lbf 5 Nm 58 4 4x 3 7 ft lbf 5 Nm 57 4 2x 3 7 ft lbf 5 Nm...
Page 133: ...page 133 of 152 59 60 61...
Page 135: ...page 135 of 152 62 63 4 4x 3 7 ft lbf 5 Nm 64...
Page 136: ...page 136 of 152 65 66 67 size 1 2x...
Page 137: ...page 137 of 152 68 69...
Page 140: ...page 140 of 152 4 638821 638821 5 6...
Page 141: ...page 141 of 152 7 T20 4x size 1 4x 8...
Page 149: ...page 149 of 152...
Page 150: ...page 150 of 152...
Page 151: ...page 151 of 152...