page 90 of 152
Attention (OUT_MODE):
If control is running, it is only possible to switch either from 1 to 2, or
from 2 to 3, or from 3 to 2. The set vacuum is adopted in each case.
Command
Operation
Response
Description
IN_SP_4
delay
XX:XX h:m
hours:minutes (00:00 = Off)
IN_SP_5
switch off
pressure
XXXX mbar/Torr/hPa
”Maximum” for ”Vac control”,
”Minimum” for ”Pump down”)
unit according to preselections
IN_SP_6
runtime
XX:XX h:m
process runtime (hours:minutes)
IN_SP_P1y
time
XX:XX:XX h:m:s
time in program step y (0......9)
(hours:minutes:seconds)
IN_SP_P2y
pressure
XXXX mbar/Torr/hPa
pressure in program step y (0......9)
unit according to preselections
IN_SP_P3y
venting valve
0
1
no venting valve in program step y
(0......9)
venting valve in program step y
(0.....9)
IN_SP_P4y
Step
0
1
no ”Step” in program step y (0.....9)
”Step” in program step y (0......9)
IN_SP_P5y
Auto
0
1
no ”Auto” in program step y (0.....9)
”Auto” in program step y (0......9)
IN_PV_Sx
current pressure
of pressure trans-
ducer x
XXXX mbar/hPa/Torr
pressure of pressure transducer x
(order of numbering according to
display in ”
Sensors
” menu)
IN_VER
version
CVC 3000 VX.XX
software version
Write commands
”
CVC 3000
”
Command
Operation
Parameter
Description
OUT_MODE
function
0
1
2
3
30
31
32
4
VACUU
•LAN
Pump down
Vac control
Auto mode
optional: sensitivity: low
optional: sensitivity: normal
optional: sensitivity: high
Program
OUT_CFG
configuration
(bus monitoring)
yXXX
X0XX
X1XX
X2XX
XX0X
XX1X
XXX0
XXX1
y: 0......D: language
+
(hexadecimal), see
”Read commands CVC 3000”
pressure unit mbar
pressure unit Torr
pressure unit hPa
Autostart off
Autostart on
acoustic signal off
acoustic signal on
Summary of Contents for MV 10C NT VARIO o.C.
Page 34: ...page 34 of 152 1 8 2 9 18 PC 3012 NT VARIO EK Peltronic 12 11 16 8 17 3 4...
Page 104: ...page 104 of 152 2 3 size 1 2x 4...
Page 105: ...page 105 of 152 6 5 7 4 4x...
Page 106: ...page 106 of 152 9 8 1 2 2 5 mm 2 5 mm 10 ME 16C NT VARIO MD 12C NT VARIO MV 10C NT VARIO...
Page 108: ...page 108 of 152 12 Emission condenser Peltronic 4 4x 11A 13 4 2x...
Page 109: ...page 109 of 152 15 14 4 4x Don t loosen this screw 15A 16...
Page 110: ...page 110 of 152 16 17 5 2x 15A 1 2 3 1 4 1 4 3 2x 5...
Page 111: ...page 111 of 152 19 20 18 5 2x...
Page 113: ...page 113 of 152 21 2 5 mm Replacing the diaphragms 22 23 696867 696839 Clean 5 8x...
Page 114: ...page 114 of 152 24 25 26 1 2...
Page 115: ...page 115 of 152 27 3 28 29 Pay attention to number of washers...
Page 118: ...page 118 of 152 36 37 38 Clean...
Page 119: ...page 119 of 152 39 40 41 IN IN EX EX 696867 696839 4x...
Page 120: ...page 120 of 152 42 43 44 IN EX T20 4x 2 2 ft lbf 3 Nm...
Page 121: ...page 121 of 152 45 46 47 Maintain all pump heads in the same way...
Page 125: ...page 125 of 152 2 size 2 2x 3 4A Version A valve 24 4 mm...
Page 126: ...page 126 of 152 5A 696839 24 4 mm 26 mm 4B 5B 696839 37 mm 40 mm Version B valve 37 mm...
Page 127: ...page 127 of 152 6 7 size 2 2x 7A ME 16C NT VARIO MD 12C NT VARIO MV 10C NT VARIO size 2 2x...
Page 129: ...page 129 of 152 49 Assembling the pump housing 50 51 5 2x 8 9 ft lbf 12 Nm...
Page 130: ...page 130 of 152 52 53 54 5 2x 8 9 ft lbf 12 Nm...
Page 131: ...page 131 of 152 55 55A 1 2 3 3 2x 55A 56...
Page 132: ...page 132 of 152 56 4 4x 3 7 ft lbf 5 Nm 58 4 4x 3 7 ft lbf 5 Nm 57 4 2x 3 7 ft lbf 5 Nm...
Page 133: ...page 133 of 152 59 60 61...
Page 135: ...page 135 of 152 62 63 4 4x 3 7 ft lbf 5 Nm 64...
Page 136: ...page 136 of 152 65 66 67 size 1 2x...
Page 137: ...page 137 of 152 68 69...
Page 140: ...page 140 of 152 4 638821 638821 5 6...
Page 141: ...page 141 of 152 7 T20 4x size 1 4x 8...
Page 149: ...page 149 of 152...
Page 150: ...page 150 of 152...
Page 151: ...page 151 of 152...