Adjustment under vacuum:

Evacuate gauge head to a 

pressure < 5 x 10

-4

 mbar

. Status indication vacuum is displayed.

Note: 

Adjustment under vacuum with an actual  pressure

 >

 5 x 10

-4

 mbar reduces the accuracy of the

measurement.

Push key-button at  the side of the housing with a pencil tip or a small screwdriver. Status indication
readjustment is displayed.

Push key-button again in order to adjust pressure reading to 1 x 10

-3

 mbar. Status indication readjust-

ment vanishes.

Note: 

If the warning triangle flashes three times, the gauge head is exposed to a pressure where

readjustment is not possible. Push key-button again to stop readjustment operation. Vent or evacuate
the vacuum system in order to adjust required pressure. If due to aging or contamination an old gauge
cannot be adjusted, replace gauge head.

Adjustment to a factory-set reference pressure:

Instead of adjustment under vacuum, adjustment to a factory-set reference pressure is possible.
The reference mark is only valid, if a 

new 

clean gauge head is used, otherwise readjustment may reduce

the accuracy of measurement. Proceed only if a pressure < 5 x 10

-4

 mbar cannot be achieved, i. e. if a

suitable vacuum pump is not available.

Note: 

If due to aging or contamination an old gauge cannot be adjusted under vacuum, replace gauge head.

Push key-button at  the side of the housing with a pencil tip or a small screwdriver. Status indication
readjustment (”Set”) is displayed.

Disconnect leads from housing connector. The warning triangle flashes.

Push key-button again in order to adjust pressure reading to the factory-set reference mark as adjust-
ment under vacuum. Status indication readjustment vanishes.

Insert leads into housing connector. The warning triangle vanishes.

Adjustment of the pressure transducer VSK 5

The device has been adjusted at the factory. In general there is no need for readjustment by the user
because of the excellent longtime stability.
Depending on operation conditions, type of application and accuracy requirements, an inspection and
readjustment may become necessary.

Adjustment at atmospheric pressure:

Admit air to the vacuum gauge. Make sure that the vacuum connection at the vacuum gauge is at
atmospheric pressure

Note: To determine the exact actual pressure, 

use an accurate barometer or get accurate reading

from the weather service, e. g. next airport..... (take into account the difference in altitude between e. g.
airport and laboratory).

Push key-button at the side of the housing by using a pencil tip or a small screwdriver. Status indication
readjustment (”Set”) and the actual pressure are displayed.

With a second tip within two seconds or continuous pressing, a pressure range from 700 to 1060 mbar
is running. Use the key-button to adjust the display to actual atmospheric pressure (short tips in steps
of 1 mbar).  When the key-button is released, the adjustment mode is terminated and the selected
pressure value is stored automatically.

Summary of Contents for VAP 5

Page 1: ...buted completely and unchanged It is strictly the users responsibility to check carefully the validity of this document with respect to his product manual no 999039 16 12 2009 Vacuum gauges VAP 5 DVR...

Page 2: ...controller VACUUBRAND vacuum gauges are the result of many years of experience in construction and practical operation of these vacuum gauges combined with the latest results in material and manufact...

Page 3: ...his product manual no 999039 16 12 2009 Attention Important notes Not permitted Misuse may cause damage Caution Hot surface Isolate equipment from mains Note Contents Safety information 4 Technical d...

Page 4: ...ination withVACUUBRAND genuine accessories Obey all other relevant safety requirements e g accident prevention regulations for working in laboratories and explosion protection guidelines and provide p...

Page 5: ...sary is approximately decontaminated before starting maintenance Use only genuine spare parts and accessories Otherwise safety and performance of the equipment as well as the electromagnetic compatibi...

Page 6: ...0 1 n i h t i w e u l a v d e t a c i d n i f o 0 1 r a b m 0 0 1 2 r a b m e g n a r t i g i d 1 r a b m 1 t n e i c i f f e o c e r u t a r e p m e T d e i f i c e p s t o n K r a b m 7 0 0 n o i t...

Page 7: ...peration Display Rear side key button for digital adjustment and electronic scale conversion pressure unit at the side of the housing male connector for mains cable thread M8 for stand rod female conn...

Page 8: ...yed with a resolution of 0 1 mbar in the range of 0 1 to 10 mbar and with a resolution of 1 mbar for pressures 10 mbar Preparing for operation Connect gauge head with connecting cable to the female co...

Page 9: ...ace the actual pressure value can be read out The single abbreviations of a command are separated by underscores ASCII 5FH The string is terminated with CR LF ASCII 0DH ASCII 0AH r e t e m a r a p e c...

Page 10: ...d a e h e g u a G d e t a n i m a t n o c r o r e t l i f e v i t c e t o r p n a e l C y r a s s e c e n f i e n o w e n e s u d n a d a e h e g u a g n a e l C t s u j d a e r g n o r w s i t n e m...

Page 11: ...h the pumped media see Technical data If the vacuum system is contaminated oil particles etc Replace the protective filter cat no 63 87 29 in the gauge head VSP 5 if necessary Using small screw driver...

Page 12: ...not be adjusted under vacuum replace gauge head Push key button at the side of the housing with a pencil tip or a small screwdriver Status indication readjustment Set is displayed Disconnect leads fro...

Page 13: ...ey button to adjust the display to the reference pressure at the vacuum line within the range of 0 10 mbar Note If the nominal ultimate vacuum of a diaphragm pump is used as reference vacuum the accur...

Page 14: ...ry product returned for repair We submit quotations only on request and always at the customer s expense If an order is given the costs incurred are offset from the costs for repair or from the pur ch...

Page 15: ...for the returned product that neither toxic corrosive biologically active explosive radio active nor contamination dangerous in any way has occurred the product is free of dangerous substances the oil...

Page 16: ...agreement of the manufacturer Par la pr sente nous d clarons que le dispositif d sign ci dessus est conforme aux prescriptions de base des directives EU applicables et indiqu s en ci que concerne con...

Page 17: ...obtained under test conditions for specific applications depend on a number of factors beyond our control It is therefore strictly the users responsibility to very carefully check the validity of app...

Reviews: