page 15 of 53
• Adopt suitable measures in case of differences from
recommended conditions, e.g., using the equipment
outdoors, installation in higher altitudes, conductive
pollution or external condensation on the pump.
• Do not operate this product near flames.
To the best of our knowledge the equipment is in com-
pliance with the requirements of the applicable EC-direc-
tives and harmonized standards (see ”Declaration of Con-
formity”) with regard to design, type and model. Directive
EN 61010-1 gives in detail the conditions under which the
equipment can be operated safely (see also IP degree of
protection, „Technical data“, pg. 23).
Operating conditions
➨
These pumps are not approved for operation in po-
tentially explosive atmospheres.
Do not operate the
pumps in potentially explosive atmospheres.
➨
Pumps
without the
”
`
”
mark on the rating plate
are
not approved
for the
pumping of potentially explo-
sive atmospheres
.
Do not pump potentially explo-
sive atmospheres
with those pumps.
➨
Pumps
bearing the
”
`
”
mark on their rating plates
are
approved for the pumping of potentially explo-
sive atmospheres
according to their classification
II 3G IIC T3 X
according to ATEX, but they are
not
approved for operation in potentially explosive at-
mospheres
Equipment marking (ATEX)“, pg. 21)
.
➨
The pumps are
not suitable
to pump any of the sub-
stances listed below.
Do not pump:
-
unstable substances
- substances which react explosively under
impact
(mechanical stress) without air
NOTICE