The reprocessing of the accessories: 

●  Breathing circuit:   

1. Dismantle all components except the stainless steel spring 

attached to the body of water trap   

2. Clean the parts with low flow running water, using a soft 

brush to remove all foreign material. (for example, soil or 

organic matter) 

3. Sterilization by autoclave: 132℃/ 4minutes. 

●  Chamber: 

1. Dismantle all components except the stainless steel spring 

attached to the body of water trap   

2. Clean the parts with low flow running water, using a soft 

brush to remove all foreign material. (for example, soil or 

organic matter) 

3. Sterilization by autoclave: 132℃/ 4minutes. 

●  Filter: 

1. Take couplings apart from the filter and use detergent 

compatible with polysulfone to clean the shell of filter with a 

damp cloth.   

2. The filter has to be sterilized before use and reuse. VADI 

only recommends steam autoclaving with wrapping the filter 

in sterilization pouch at temperature at 132℃  for 4 minutes 

(require at least 20 minutes drying cycle after sterilization), 

complying with the steam autoclave manufacturer’s 

instructions . 

 

 

Summary of Contents for VH-1500

Page 1: ...VH 1500 RESPIRATORY HUMIDIFIER Manual Read instruction carefully before use Made In Taiwan VA 1 800 VH1500 110 EN_01...

Page 2: ...er 4 Setting up the humidifier 4 Setting up the humidifier 4 Setting up the humidifier 6 6 6 6 8 8 8 8 5 Performance 5 Performance 5 Performance 5 Performance Data Data Data Data 9 9 9 9 6 Disposition...

Page 3: ...repair the device by yourself if it is malfunction Please contact local distributor for the service Must be used by qualified medical staff or trained personnel To avoid the risk of electric shock th...

Page 4: ...er trap 2 Clean the parts with low flow running water using a soft brush to remove all foreign material for example soil or organic matter 3 Sterilization by autoclave 132 4minutes Filter 1 Take coupl...

Page 5: ...d Compliance EN IEC 60601 1 EN 60601 1 2 Classified as Class I Type BF IPX1 Definitions of symbols Definitions of symbols Definitions of symbols Definitions of symbols TYPE B TYPE B TYPE B TYPE BF F F...

Page 6: ...t length 22mmID 120cm VADI chamber G 314004 Power Level Heater plate Temperature Flow rate range LPM Temperature at patient Y 3 52 5 40 26 30 5 60 5 40 30 34 8 71 5 40 33 36 Temperature accuracy 3 Ove...

Page 7: ...is button to cut off the alarm signal alarm continues signal after 2 minutes The scheme of power level shows the temperature of heater plate Caution Heater level is Stand by heater plate is not heated...

Page 8: ...heater base Pic 2 Caution Choose the appropriate chamber 4 3 Add distilled water into chamber Pic 3 Or stab distilled water with the needle of IV SET and connect the tube to humidification chamber Pi...

Page 9: ...in chamber may result water condensation in breathing circuit Do not touch the heater plate while operation Do not use any blanket to cover the breathing circuit Operating environment Suggested enviro...

Page 10: ...erved during normal operation If there is no flow turn off the humidifier Water traps are recommended for use to collect water condensation due to differences in temperature Absorbent paper is recomme...

Page 11: ...check if the temperature is correspondent to the indicator on device panel Suggested Tester Fluke 572 Thermometer Accuracy 1 0 Note Please make performance test every 12 months 6 Disposition Please di...

Page 12: ...sable adult humidification chamber Reusable adult humidification chamber Reusable adult humidification chamber Reusable adult humidification chamber G G G G 314005 314005 314005 314005 Reusable infant...

Reviews: