The reprocessing of the accessories:
● Breathing circuit:
1. Dismantle all components except the stainless steel spring
attached to the body of water trap
2. Clean the parts with low flow running water, using a soft
brush to remove all foreign material. (for example, soil or
organic matter)
3. Sterilization by autoclave: 132℃/ 4minutes.
● Chamber:
1. Dismantle all components except the stainless steel spring
attached to the body of water trap
2. Clean the parts with low flow running water, using a soft
brush to remove all foreign material. (for example, soil or
organic matter)
3. Sterilization by autoclave: 132℃/ 4minutes.
● Filter:
1. Take couplings apart from the filter and use detergent
compatible with polysulfone to clean the shell of filter with a
damp cloth.
2. The filter has to be sterilized before use and reuse. VADI
only recommends steam autoclaving with wrapping the filter
in sterilization pouch at temperature at 132℃ for 4 minutes
(require at least 20 minutes drying cycle after sterilization),
complying with the steam autoclave manufacturer’s
instructions .