background image

- 3 - 

2. Product Specification

 

Type:VH-1500 

Dimensions:145mm(H)*170mm(D)*135mm(W)                   

Weight:1.1kg 

Electrical Ratings 

Supply voltage:100-120Vac ~1.6A or 200-240Vac ~0.8A 

                             

(Please refer to the mark on device.)

 

Supply Frequency:50/60Hz   

Sine Wave Max. power supply: 180W 

Device disconnection:pull out the power cord of this device from 

socket. 

Max. operating pressure:20kpa 

Warm-up time:less than 30min. 

This device life cycle is 3 years. 

Transport condition:-10 to +55℃ 

(Humidity RH: 5%~95%,without any water condensation)   

Storage condition:-20 to +65℃ 

(Humidity RH: 5%~95%,without any water condensation)   

Suggested operating environment temperature: 18~26°. 

Suggested operating environmental pressure: 700~1060 hPa. 

Humidity Output: >33mg/l within recommended flow range. 

Standard Compliance:EN/IEC-60601-1,EN-60601-1-2 

Classified as:Class I, Type BF, IPX1 

Definitions of symbols

Definitions of symbols

Definitions of symbols

Definitions of symbols

 

 

TYPE B

TYPE B

TYPE B

TYPE BFFFF    Applied Part

Applied Part

Applied Part

Applied Part    

(Inhalation tube)

(Inhalation tube)

(Inhalation tube)

(Inhalation tube)    

    

Consult operating 

Consult operating 

Consult operating 

Consult operating      

instructions

instructions

instructions

instructions    

    

CAUTION

CAUTION

CAUTION

CAUTION    

Hot surface may

Hot surface may

Hot surface may

Hot surface may    

exceed 75

exceed 75

exceed 75

exceed 75℃

℃    

 

Drip Proof

Drip Proof

Drip Proof

Drip Proof    

 

Caution

Caution

Caution

Caution    

Electrical Shock Hazard. 

Electrical Shock Hazard. 

Electrical Shock Hazard. 

Electrical Shock Hazard.      

Refer to qualified service 

Refer to qualified service 

Refer to qualified service 

Refer to qualified service      

Personnel.

Personnel.

Personnel.

Personnel.    

 

CE Mark

CE Mark

CE Mark

CE Mark    

Summary of Contents for VH-1500

Page 1: ...VH 1500 RESPIRATORY HUMIDIFIER Manual Read instruction carefully before use Made In Taiwan VA 1 800 VH1500 110 EN_01...

Page 2: ...er 4 Setting up the humidifier 4 Setting up the humidifier 4 Setting up the humidifier 6 6 6 6 8 8 8 8 5 Performance 5 Performance 5 Performance 5 Performance Data Data Data Data 9 9 9 9 6 Disposition...

Page 3: ...repair the device by yourself if it is malfunction Please contact local distributor for the service Must be used by qualified medical staff or trained personnel To avoid the risk of electric shock th...

Page 4: ...er trap 2 Clean the parts with low flow running water using a soft brush to remove all foreign material for example soil or organic matter 3 Sterilization by autoclave 132 4minutes Filter 1 Take coupl...

Page 5: ...d Compliance EN IEC 60601 1 EN 60601 1 2 Classified as Class I Type BF IPX1 Definitions of symbols Definitions of symbols Definitions of symbols Definitions of symbols TYPE B TYPE B TYPE B TYPE BF F F...

Page 6: ...t length 22mmID 120cm VADI chamber G 314004 Power Level Heater plate Temperature Flow rate range LPM Temperature at patient Y 3 52 5 40 26 30 5 60 5 40 30 34 8 71 5 40 33 36 Temperature accuracy 3 Ove...

Page 7: ...is button to cut off the alarm signal alarm continues signal after 2 minutes The scheme of power level shows the temperature of heater plate Caution Heater level is Stand by heater plate is not heated...

Page 8: ...heater base Pic 2 Caution Choose the appropriate chamber 4 3 Add distilled water into chamber Pic 3 Or stab distilled water with the needle of IV SET and connect the tube to humidification chamber Pi...

Page 9: ...in chamber may result water condensation in breathing circuit Do not touch the heater plate while operation Do not use any blanket to cover the breathing circuit Operating environment Suggested enviro...

Page 10: ...erved during normal operation If there is no flow turn off the humidifier Water traps are recommended for use to collect water condensation due to differences in temperature Absorbent paper is recomme...

Page 11: ...check if the temperature is correspondent to the indicator on device panel Suggested Tester Fluke 572 Thermometer Accuracy 1 0 Note Please make performance test every 12 months 6 Disposition Please di...

Page 12: ...sable adult humidification chamber Reusable adult humidification chamber Reusable adult humidification chamber Reusable adult humidification chamber G G G G 314005 314005 314005 314005 Reusable infant...

Reviews: