Heater control:
Heater control:
Heater control:
Heater control:
settings from 0 to 9. Power Level settings:
0
1
2
3
4
5
6
7
8
9
Stand by 45℃ 48℃ 52℃ 56℃ 60℃ 64℃ 68℃ 71℃ 75℃
Max. Flow : 40 LPM
Max. Delievered Gas Temp. : 37.9℃ @15LPM
(Power Level : 9, room temperature 23℃± 2℃, 120cm-silicone circuit)
Performance :
Performance :
Performance :
Performance :
Temperature difference between heater plate and patient’s side
(room temperature 23℃± 2℃, silicone circuit length 22mmID / 120cm, VADI
chamber G-314004):
Power Level
Heater plate
Temperature (℃)
Flow rate range
(LPM)
Temperature at
patient Y(℃)
3
52
5~40
26~30
5
60
5~40
30~34
8
71
5~40
33~36
Temperature accuracy:±3℃。
Over-heated protection:
if the heater temperature exceeds 95±6℃, the power to the
heated plate is discontinued and the Power-on/off will not be
operative.
Low Temperature Alarm:
30 min after operation, the temperature of the heater plate is 5℃
lower than setting. The display will be shown “L” and emit an
alarm.
High Temperature Alarm:
While the temperature of heater plate is over 85±3℃, the display
will be shown “H” and emit an alarm.