1. Establish a connection to AQT530.
2. Adjust the gain and offset of the parameters.
For example, to adjust offset up by 100 ppb and sensitivity 1.2 times higher, type:
set
o3_zero=100
set
o3_span=120
3. To save the changes:
write
--really
The new values overwrite the previous zero and span values.
Values without linear correction are available with Modbus registers 0x006C - 0x0075.
More information
‣
Changing configuration parameter (set command) (page 35)
‣
Saving to non-volatile memory (write command) (page 42)
Chapter 4 – CLI operation
27
Summary of Contents for AQT530
Page 1: ...M212572EN C Configuration Guide Vaisala Air Quality Transmitter AQT530...
Page 62: ......
Page 63: ......
Page 64: ...www vaisala com...