5.9.2 Sending HMP155 for Calibration
Once a year, or if the sensor readings seem inaccurate, send HMP155 to Vaisala for
calibration. Replace it with a calibrated spare probe, if available, or wait until Vaisala sends
the probe back from calibration.
1. Remove the probe from inside the radiation shield.
2. Send the probe to Vaisala for calibration.
3. Check the operation of the calibrated probe by warming the sensor head with your
hand, and monitor the value change.
4. Place the calibrated probe inside the radiation shield.
5.9.3 Adding HMP155 to System
Before configuring the new sensor, complete the mechanical installation.
1. In AWS Client, select
Settings > Parameters > Static
.
2. In the
CFG
parameter set, select
HMP155
(serial sensor) or
HMP155_ANL
(analog
sensor).
3. To enable the sensor, type
1
in the
Value
field, and select
Save
.
4. In the
RH_Selector
parameter set, select the relative humidity sensor to use.
•
1
= HMP155 (serial)
•
2
= HMP155 (analog)
5. In the
TA_Selector
parameter set, select the air temperature sensor to use.
•
1
= HMP155 (serial)
•
2
= HMP155 (analog)
6. To activate the new configuration, reset the QML data logger. Select
Maintenance >
Reset > Immediate
.
5.10 RG13(H) Maintenance
To ensure accurate operation, keep the sensor clean. Clean the sensor at least every
6 months. When leaves are falling and there is a lot of debris in circulation, check the sensor
more often.
To avoid erroneous measurements, avoid tipping the cup assembly when the gauge is
connected to the weather station and the station is operating.
To ensure reliable and accurate measurements, Vaisala recommends you to carry out the
following checks at each visit to the rain gauge.
1. Inspect the funnel and filter for any damage or blockage.
2. Remove any obstacles from the funnel, such as leaves, debris, dirt, and dust.
3. Clean the filter by removing the end cap from the filter tube.
AWS310-SITE Configuration and Maintenance Manual
M211828EN-C
106
Summary of Contents for AWS310-SITE
Page 10: ...AWS310 SITE Configuration and Maintenance Manual M211828EN C 8 ...
Page 21: ...Chapter 2 Product Overview 19 ...
Page 22: ...AWS310 SITE Configuration and Maintenance Manual M211828EN C 20 ...
Page 62: ...AWS310 SITE Configuration and Maintenance Manual M211828EN C 60 ...
Page 86: ...AWS310 SITE Configuration and Maintenance Manual M211828EN C 84 ...
Page 140: ...AWS310 SITE Configuration and Maintenance Manual M211828EN C 138 ...
Page 148: ...AWS310 SITE Configuration and Maintenance Manual M211828EN C 146 ...
Page 150: ...AWS310 SITE Configuration and Maintenance Manual M211828EN C 148 ...
Page 152: ...AWS310 SITE Configuration and Maintenance Manual M211828EN C 150 ...
Page 166: ...AWS310 SITE Configuration and Maintenance Manual M211828EN C 164 ...
Page 172: ...AWS310 SITE Configuration and Maintenance Manual M211828EN C 170 ...
Page 174: ...AWS310 SITE Configuration and Maintenance Manual M211828EN C 172 ...
Page 175: ......
Page 176: ...www vaisala com ...