9.2 Component specifications
9.2.1 Vaisala Differential Pressure Transmitter PDT101
specifications
Table 10 PDT101 measurement performance
Property
Description/Value
Measurement ranges (bidirectional)
±60 Pa
±0.25 in H
2
O
Accuracy (incl. non-linearity, hysteresis, repeatability
and zero/span calibration settings)
0.4 % span
Long-term stability
≤
0.5 % span/year
Response time (10 ... 90 %)
250 ms
Warm-up time
15 s
Compensated temperature range
+2 ... +54 °C
(+35.6 ... +129.2 °F)
Temperature dependence
±(0.065 Pa + 0.054 % of reading) / °C
or ±(0.00015 in H
2
O + 0.03 % of reading) / °F
(reference 21 °C or 70 °F)
Pressure type
Differential, gauge, vacuum and compound