Chapter 5 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 39
Stop Measurement Output
You can stop the measurement output by pressing
ESC
, or with the
S
command:
S
<cr>
Set Output Interval
Use the
INTV
command to change the output interval of the
automatically repeating measurement messages. The measurement
messages are repeated in the RUN mode, or after the
R
command has
been given.
INTV
[
n xxx
]<cr>
where
n
= time interval, range 0 ... 255
xxx
= time unit = "S", "MIN", or "H"
The shortest output interval (with n = 0) outputs the measurement
messages as quickly as the transmitter’s internal measurement cycle
produces them, without additional delay.
Example:
intv 1 min
Output interval: 1 MIN