USER'S GUIDE____________________________________________________________________
54 __________________________________________________________________ M211371EN-B
Error Messages
The transmitter has error codes that can be displayed on the serial line
using the
ERRS
command. See section Show Currently Active Errors
on page 37.
Table 10
Error Messages
Error Message
Possible Cause and Solution
T MEAS error
F meas error
MPS1 sensor damaged or missing.
Contact a Vaisala service center.
Voltage error
Voltage too low error
Supply voltage out of range.
Check and correct.
Program flash check sum error
Parameter flash check sum error
INFOA check sum error
SCOEFS check sum error
CURRENT check sum error
Internal transmitter error. If the error
remains after a reset and restoring the
factory settings, contact a Vaisala
service center.
Unknown Serial Settings
If you are not sure what the current serial settings of your transmitter
are, follow the steps below to connect to the transmitter. This
procedure assumes you have the USB service cable available.
For more information on the USB service cable and the PuTTY
terminal application, see section Serial Communication on page 32.
1.
Connect the USB service cable to your PC, but do not connect it
to the transmitter yet.
2.
Start a terminal application and open a terminal session using
the default serial settings of 19200 8 N 1.
3.
Plug the USB service cable into the transmitter, and
immediately send the hash character # to the serial line:
#
4.
The normal startup of the transmitter is now interrupted, and the
transmitter is available for commands with the default serial
settings.
5.
You can now check the transmitter’s information by issuing
the
?
command:
?
The transmitter’s currently stored serial settings are visible on
the line marked
Baud P D S
.