User's Guide _______________________________________________________________________
34 ___________________________________________________________________ M210397EN-D
0610-073
Figure 19
DMT142 Dewpoint Measurement Accuracy Graph
Table 8
Operating Environment Specifications
Property Description
/
Value
Temperature
1)
0 … +60 °C (+32 ... +140 °F)
Relative humidity
0 ... 100 %RH
Pressure
1)
0 ... 50 bar
a
(0 ... 725 psi
a
)
Sample flow rate
no effect on measurement accuracy
Measured gases
noncorrosive gases
1) For extended temperature down to -40 °C (-40 °F) or pressure up to 50 bar
a
(725 psi
a
) the
supply voltage must be 24 … 28 VDC
Table 9
Output Specifications
Property Description
/
Value
Analog output (scalable)
4 ... 20 mA (three-wire)
0 ... 1 V
0 ... 5 V
Resolution for current output
0.002 mA
Resolution for voltage output
0.3 mV
Typical temperature dependance
0.005 % of span/°C
Connector
4-pin M8 (IEC 60947-5-2)
AERO AE65-221-FA35PN (DMT142
PALL OEM model)
2 m connection cable with snap-on
or thread locking available
RS232 serial line for service use
with DMT142RS cable
Summary of Contents for Drycap DMT142
Page 1: ...USER S GUIDE Vaisala DRYCAP Dewpoint Transmitter DMT142 M210397EN D...
Page 48: ......
Page 49: ......
Page 50: ...www vaisala com M210397EN...