Vaisala GMP 251/252 CO

Probe 

Quick Start Guide 

 

 

The Vaisala GMP251 and GMP252 Carbon Dioxide probes are compatible with the 
Nexsens X2 data logger using RS485 Modbus RTU output.  

Note that the GMP25x sensor membrane and cable connection are NOT 
waterproof.

 A waterproof, gas-permeable membrane must be installed on the 

sensor and the green M12 cable connection must be waterproofed prior to 
deployment. 

 

 

 

Wiring 

Note: 

Most sensors

 

will have factory-installed UW connectors. If this is the case, 

skip to the 

Configuring the Vaisala GMP 251/252 

section. 

For compatibility, a Vaisala sensor cable must be fitted with a NexSens UW 
connector to interface with the X2 data logger.  

For user-wiring with a NexSens UW-Field Wireable Plug (UW-FWP) follow the wiring 
diagram below. 

Vaisala 223263SP Cable 

Wire Color 

Sensor

 

Signal 

UW-FWP J Pin 

Connection 

Brown 

12 to 30 VDC Power 

J5 

Blue 

Power GND 

J2 

White 

RS485-B 

J7 

Black 

RS485-A 

J8 

 
Configuring the Vaisala GMP 251/252 

Note: 

All GMP25x sensors not integrated by NexSens technology require the use of 

a PC terminal connection and Vaisala’s USB interface cable to modify the sensor 
output.

 

1.

 

Download or launch a terminal emulation software. NexSens recommends 

Tera Term

 as a free open-source option.  

 

2.

 

Connect the Vaisala USB cable and verify Windows properly downloads the 
necessary drivers automatically. Verify this appears as a unique COM port 