Close the Connection in POLL Mode
The
CLOSE
command closes the connection to the probe.
CLOSE
<cr>
Example:
close
line closed
Reset the Probe
Use the
RESET
command to reset the probe. Upon reset or power-up,
the probe enters the serial mode that has been set with the
SMODE
command.
RESET
<cr>
Example (probe set to serial mode STOP, will output probe model and
software version at reset):
reset
HMP60 / 1.00.0
Example (probe set to serial mode RUN, will start to output measurement
messages at reset):
reset
T= 23.6 'C RH= 20.2 %RH Td= -0.5 'C
T= 23.6 'C RH= 20.2 %RH Td= -0.5 'C
T= 23.3 'C RH= 20.2 %RH Td= -0.8 'C
...
Restore Factory Settings
Use the
FRESTORE
command to restore the factory settings to the
probe. All user settings, including the user-performed calibration
corrections, will be lost. The probe will revert back to the factory
calibrated settings.
FRESTORE
<cr>
Example:
frestore
Factory settings restored