Chapter 4 _________________________________________________________________ Operation
VAISALA ________________________________________________________________________ 39
List of Serial Commands
All commands can be issued either in uppercase or lowercase. In the
command examples, the keyboard input by the user is in
bold
type.
The notation <cr> refers to pressing the carriage return (
Enter
) key on
your computer keyboard. Press
Esc
to clear the command buffer before
starting to enter commands.
Table 11
List of Serial Commands (software version 2.0.7)
Command
Description
?
Output information about the device
AERR
Set analog output error level
AMODE
View or set the analog output mode
AOVER [ON/OFF]
Allow analog outputs to exceed their range
10%
ASEL
Set analog output parameters and scaling
CDATE
View or set the calibration date
CODE
View the order code of the probe
CRH
Calibrate and adjust RH measurement
CRHCLR
Clear adjustment of RH measurement
CT
Calibrate and adjust T measurement
CTCLR
Clear adjustment of T measurement
CTEXT
View or set the calibration information field
ERRS
List present probe errors
FILT [0.001 … 1]
Set the result filtering
FRESTORE
Restore factory settings
HELP
List available commands
INTV [0 ... 255 S/MIN/H]
Set the continuous output interval (for RUN
mode)
L
Displays user adjustment parameters
R
Start the continuous outputting
RESET
Reset the probe
RHLIMIT
Extend maximum RH reading
S
Stop the continuous outputting
SDELAY [0 … 255]
View or set serial line answer minimum
delay
SEND [0 ... 255]
Output the reading once
SNUM
View the serial number of the probe
UNIT
Select metric or non-metric output units
VERS
View software version of the probe