User's Guide _______________________________________________________________________
40 ___________________________________________________________________ M211060EN-H
Table 12
Additional Commands for Probes with RS-485 Output
Command
Description
??
Output information about the device in
POLL mode
ADDR [0 ... 255]
Set the probe address (for POLL mode)
CLOSE
Close the temporary connection (Back to
POLL mode)
OPEN [0 ... 255]
Open a temporary connection to a POLL
mode device
SERI [baud p d s]
User Port settings (Default: 19200 N 8 1)
baud: 300 ... 57600
SMODE
[STOP/RUN/POLL/MODBUS/
VDIGI/ANALOG]
Set the serial interface mode
Device Information and Status
View Device Information
The
?
command outputs a listing of device information.
?
<cr>
Example (output from HMP63):
?
HMP63 / 1.0.4
Serial number : H3640004
Batch number : T0001109
Sensor number : H0000322
Sensor model : Intercap
Order code : A12A0A2B0
Cal. date : 20120907
Cal. info : VAISALA/HEL
Time : 00:21:05
Serial mode : ANALOG
Baud P D S : 19200 N 8 1
Output interval: 1 S
Serial delay : 30
Analog delay : 10 S
Address : 0
Filter : 1.000
Ch1 output : 0 ... 1 V
Ch2 output : 0 ... 1 V
Ch1 RH lo : 0.00 %RH
Ch1 RH hi : 100.00 %RH
Ch2 T lo : -40.00 'C
Ch2 T hi : 60.00 'C
If the probe is in poll mode, but a connection has not been opened using
the
OPEN
command, issue the
??
command. For a description of the
serial interface modes, see section Set Serial Line Settings on page 43.
??
<cr>