5.
Connect the jumper to the RH offset pins (see Figure 5 below) and
adjust the output signal with push buttons to the value given in the
calibration table (Table 5 on page 25).
6.
Insert the probe into the measurement hole of the NaCl chamber.
7.
Wait for 20 to 40 minutes.
8.
Check that the reading corresponds within the desired accuracy to
the reading given in the calibration table. If not, connect the jumper
to the RH gain pins and adjust the reading with push buttons.
9.
Check again the reading at the first point and repeat the procedure
if necessary.
T offset
RH gain (WET)
T gain
RH offset (DRY)
Figure 5
Calibration Pins and Pushbuttons
With the jumper connected to the RH gain pins, each pressing of the UP
or DOWN button changes the value approximately 0.05 %RH at
75 %RH. With the jumper connected to the RH offset pins, each pressing
changes the value approximately 0.05 %RH.
With the jumper connected to the T offset pins, each pressing of the UP
or DOWN button changes the value 0.05 °C or 0.05 °F in all
temperatures. With the jumper connected to the T gain pins, each
pressing changes the value 0.05 °C or 0.05 °F in the temperature
currently prevailing in the process.