7.5 TMP1 Specifications
Table 24 Measurement Performance
Property
Description/Value
Sensor
Pt100 RTD Class F0.1 IEC 60751
Measurement range
-70 ... +180 °C (-94 ... +356 °F)
Accuracy at +23 °C (+73.4 °F)
±0.1 °C (±0.18 °F)
1)
Defined against calibration reference
2) Exposing temperature sensor to temperatures outside -20 ... +150 °C (-4 ... +302 °F) may
cause permanent additional deviation of ±0.1 °C (0.18 °F)
0
0.1
0.2
0.3
0.4
A
c
c
u
ra
c
y
[
+
/-
°
C]
Temperature [
°
C]
-4
0
-3
0
-2
0
-1
0
0
10
2
0
3
0
4
0
5
0
6
0
7
0
8
0
9
0
10
0
11
0
12
0
13
0
14
0
150
160
17
0
18
0
0.5
-70
-6
0
-50
Figure 26 TMP1 Temperature Measurement Accuracy over Full Range (Including Non-Linearity and
Repeatability)
Table 25 SI Traceable Calibration
Property
Description/Value
Uncertainty of temperature calibration (
k
= 2)
±0.1 °C (±0.18 °F) at +23 °C (+73.4 °F)
Table 26 Operating Environment
Property
Description/Value
Operating temperature range for probe body
-40 ... +80 °C (-40 ... +176 °F)
Operating temperature range for probe head
-70 ... +180 °C (-94 ... +356 °F)
Chapter 7 – Technical Data