Chapter 4 ________________________________________________________________ Operation
VAISALA____________________________________________________________________ 51
AQTEST Test Analog Outputs for
Desired Readings
Syntax: AQTEST x yyy.yyy<cr>
Use command AQTEST to test current values. Current output is
forced to correspond the chosen values.
where
x
= Output quantity of analog channel (use
abbreviations, see FORM Serial Output Format on
page 39)
yyy.yyy =
Value
Example:
>aqtest td 30<cr>
CH1 Td : 30.0000 'C 12.00000mA
CH2 x : 2.5304 g/kg 4.08097mA
>
The set values remain valid until you give the command AQTEST
without a value or reset the transmitter.
AERR Set Error Outputs
Syntax: AERR<cr>
Factory default state for analog outputs during error condition is
0 mA. Please be careful when selecting the new error value, the error
state of the transmitter should not cause problems in process
monitoring.
Example:
>aerr<cr>
Ch1 error out : 0.000mA ?
Ch2 error out : 0.000mA ?
NOTE
Error output value needs to be within a valid range for output type.