USER'S GUIDE____________________________________________________________________
70 _______________________________________________________________ M210619EN-C
6.
When the reading is stabilized, type the reference temperature
after the question mark and press
Enter
three times.
>ct
T : 16.06 Ref1 ? c
T : 16.06 Ref1 ? c
T : 16.06 Ref1 ? c
T : 16.06 Ref1 ? c
T : 16.06 Ref1 ? c
T : 16.06 Ref1 ? 16.0
Press any key when ready ...
T : 16.06 Ref2 ?
OK
>
OK indicates that the calibration has succeeded.
7.
Take the probe out of the reference conditions and replace the
filter. When using the stainless steel filter, take care to tighten
the filter properly (recommended force 130 Ncm).
8.
If needed, record the calibration information (text and date) to
the memory of the transmitter, see see CTEXT Set Calibration
Information Text on page 65 and CDATE Set Calibration Date
on page 66.
9.
Reset the transmitter by issuing the RESET command. The
transmitter returns to normal mode.