Chapter 8 ____________________________________________________________ Technical Data
VAISALA____________________________________________________________________ 73
Response time (90 %) for
HUMICAP®180R and
HUMICAP®180RC at 20 °C in
0.1 m/s air flow
17 s with grid filter
50 s with grid + steel netting filter
60 s with sintered filter
Table 12
Temperature Specifications (+ Operating Pressure
Ranges)
Property
Description / Value
HMT311
-40 ... +60 °C (-40 ... +140 °F)
HMT313 80 °C
-40 ... +80 °C (-40 ... +176 °F)
HMT313 120 °C
-40 ... +120 °C (-40 ... +248 °F)
HMT314
-70 ... +180 °C (-94 ... +356 °F),
0 ... 10 MPa (0 ... 100 bar)
HMT315 (vapor tight)
-70 ... +180 °C (-94 ... +356 °F)
HMT317 (vapor tight)
-70 ... +180 °C (-94 ... +356 °F)
HMT318
-70 ... +180 °C (-94 ... +356 °F),
0 ... 4 MPa (0 ... 40 bar)
Accuracy at +20 °C (+68 °F)
±0.2 °C
Accuracy over temperature range
See graph below
Temperature sensor
Pt 100 RTD 1/3 Class B IEC 751
∆
°C
°C
-0.7
0.7
0.6
0.5
0.4
0.3
0.2
0.1
-0.1
-0.2
-0.3
-0.4
-0.5
-0.6
0
0
-60
-40
-20
20
40
60
80
100
120
140
160
180
-80
0507-021
Figure 18
Accuracy over Temperature Range