12 Ethernet connection specification
203
Opening the instrument homepage:
1. Establish a working Ethernet connection (see above) to the DTR.
2. Start your preferred web browser (for example Internet Explorer, Mozilla Firefox,
Safari, Opera or Chrome).
3. The address (URL) of the instrument homepage is the DTR’s IP address, for a fac-
tory set DTR it is http://192.168.23.254/ (the figure 12.10 above uses a different
address, not the default). Give the address to the browser just like you’d enter any
other address (for example http://www.vaisala.com/)
4. Wait until the homepage is loaded, this may take a few seconds.
If the page looks strange in the first try, refresh/reload the page and it should settle
to look approximately like in Figure 12.10; the exact look of the page depends on
your browser and screen settings so slight variation can be expected.
5. Use the links in the link bar on the left side of the page to find more extensive
information on the instrument.
12.4.1 Remote panel
Figure 12.11
DTR remote panel
The instrument remote panel is a fully
functional virtual DTR where keys on
the keyboard are clicked with a mouse.
The DTR doesn’t make any difference be-
tween commands coming from the ac-
tual keyboard and from a remote panel.
All commands are executed in the or-
der the DTR gets them, independent of
where they come from.
Note:
The DTR display picture on the
remote panel sometimes has a slight
lag (of few seconds) before it refreshes.
This depends on many factors like the
computer and network used. If the DTR
seems to ’skip’ displays, it may be that it
is executing mouse-click keyboard com-
mands faster than your browser updates the picture.
12.4.2 Sensor verification certificate
A sensor verification certificate can be viewed and printed by following the Verifica-
tion link on the link bar. For more information on instrument verification, see Chap-
ter 13.
Summary of Contents for K-PATENTS PR-23 AX Series
Page 1: ...IM EN PR23 v 2 00 Instruction Manual Vaisala K PATENTS Process Refractometer PR 23...
Page 2: ......
Page 4: ......
Page 17: ...2 Inline refractometer sensor 7 2 2 2 PR 23 mounting guide...
Page 26: ...16 PR 23 instruction manual...
Page 29: ...4 Prism wash systems 19 Figure 4 1 A prism wash system for steam non sanitary...
Page 30: ...20 PR 23 instruction manual Figure 4 2 A sanitary prism wash system for steam...
Page 32: ...22 PR 23 instruction manual Figure 4 5 Wiring for a prism wash system for steam...
Page 33: ...4 Prism wash systems 23 Figure 4 6 A prism wash system for high pressure water non sanitary...
Page 34: ...24 PR 23 instruction manual Figure 4 7 A sanitary prism wash system for high pressure water...
Page 35: ...4 Prism wash systems 25 Figure 4 8 Wiring for a prism wash system for high pressure water...
Page 40: ...30 PR 23 instruction manual...
Page 50: ...40 PR 23 instruction manual...
Page 72: ...62 PR 23 instruction manual...
Page 174: ...164 PR 23 instruction manual...
Page 179: ...11 Safe Drive 169 11 3 3 Safe Drive steam wash system parts...
Page 186: ...176 PR 23 instruction manual 11 4 2 Wiring Figure 11 9 PR 23 SD system wiring...
Page 204: ...194 PR 23 instruction manual Figure 11 19 Markings on the retractor handle...
Page 206: ...196 PR 23 instruction manual...
Page 218: ...208 PR 23 instruction manual...
Page 226: ...216 PR 23 instruction manual...
Page 230: ...220 PR 23 instruction manual...
Page 232: ...222...
Page 238: ...228...
Page 240: ...230...
Page 242: ...232...
Page 244: ...234...
Page 246: ...www vaisala com...