58
MI instruction manual
Document/Revision No.
IM-EN-MI
/2.00
Effective: September 2, 2019
Finish by tapping here
to save settings
Tap for output
test page
Output parameters
Clears all settings
for this output
Figure 6.11
mA output settings
Important:
When you have chosen the settings for the outputs, tap
DONE
to save
changes and exit.
6.5 General I/O module PR-50024
The General I/O module card adds digital inputs and outputs to Multichannel user in-
terface MI. The module card provides 8 input/output channels.
The input and output terminals are galvanically isolated according to
•
UL 2500 V rms (UL1577)
•
CSA 800 V rms (1131V peak) maximum working voltage - basic insulation
•
CSA 400 V rms (566V peak) maximum working voltage - reinforced insulation
•
VDE Reinforced insulation 560V peak
•
TüV Reinforced insulation 400V rms maximum working voltage
Cable specifications
Conductor cross section
0.14-1.5 mm
2
(28-16 AWG)
Tightening torque, max.
0.25 Nm (0.18 lbf.ft)
Summary of Contents for MI
Page 1: ...IM EN MI v 2 00 Instruction Manual Multichannel User Interface MI ...
Page 4: ...Document Revision No IM EN MI 2 00 Effective September 2 2019 ...
Page 8: ...MI instruction manual Document Revision No IM EN MI 2 00 Effective September 2 2019 ...
Page 11: ...2 Mounting 3 Copyright Vaisala 2019 All rights reserved Figure 2 2 Enclosure dimensions ...
Page 83: ...8 Specifications 75 Copyright Vaisala 2019 All rights reserved 8 2 Parts ...
Page 84: ...76 MI instruction manual Document Revision No IM EN MI 2 00 Effective September 2 2019 ...
Page 86: ...78 MI instruction manual Document Revision No IM EN MI 2 00 Effective September 2 2019 ...
Page 88: ...80 MI instruction manual Document Revision No IM EN MI 2 00 Effective September 2 2019 ...
Page 90: ...82 MI instruction manual Document Revision No IM EN MI 2 00 Effective September 2 2019 ...
Page 92: ...www vaisala com ...