PTB 200
25 (28)
Digital Barometers
SSD/Operating Manual
24th February 1993
PTB200-O0284-1.1
First precalibrate the barometer at up to eight pressure levels. Then
calculate the corrections (reference value minus PTB 200 barometer
indication) at each pressure level and enter the corrections using the
MPCI commands.
>MPCI<cr>
1. PTB reading ?499.64<cr>
correction ?0.024<cr>
2. PTB reading ?599.79<cr>
correction ?0.013<cr>
...
7. PTB reading ?1100.33<cr>
correction ?0.027<cr>
8. PTB reading ?<cr>
>MPC ON <cr>
Multip. corr. : ON
Reading
Corrections
499.6400
0.0240
599.7900
0.0130
.
.
.
.
.
.
1100.3300
0.0270
Use ESC to abort without executing a command.
Return the MEMORY jumper to the EEPROM write disable position.