>txci
TXD0
: -1.199999971E-01 ? -0.10<cr>
(
offset
)
TXD1
:
1.00000 ? 0.99779<cr>
(gain)
TXD2
:
0.000000000E-01 ?<cr>
(factory use only)
>
The factory setting TXD2 is passed by pressing enter. Use ESC to
abort without executing the command.
Reset the transmitter to activate the given corrections.
NOTE
Always remember to return the switch SW4 to write DISABLE
position (OFF).
V0XCI Entering corrections for HMP45A-P temperature
measurement
V0XCI x
where:
x = 2 if there is one pressure transducer installed
3 if there is two pressure transducers installed
The V0XCI command is used to enter new offset and offset/gain
corrections for temperature measurement of HMP45A-P.
Entering new corrections is protected by switch SW4. Turn it to write
ENABLE position (ON) to be able to enter new offset/gain
corrections.
The factory setting V0XD2 is passed by pressing enter. Use ESC to
abort without executing the command.
Reset the transmitter to activate the given corrections.
In the following, an example of performing an offset and gain
adjustment:
>V0XCI 2
V0XD0
:
0.000000000E-01 ? -0.10<cr>
V0XD1
:
1.00000 ? 0.99778<cr>
V0XD2
:
0.000000000E-01 ?<cr>
>
NOTE
Always remember to return the switch SW4 to write DISABLE
position (OFF).