11. Select menu item
[5] Note
to edit the calibration info text that is stored in the probe. Edit
the text using the select button and arrow keys. When done, select the
OK
character in
the bottom right corner to save the changed text. To exit without saving, press the
Cancel
button.
12. Select menu item
[6] Apply
to view the calibration result. Verify the applied corrections in
the confirmation screen and press the
Apply
button to apply the adjustment to the probe,
or
Cancel
to exit without applying the adjustment.
13. Disconnect the probe from the HM40.
14. Connect the probe to the RFL100 Data Logger. See
Connecting the Probe (page 29)