9
2.3 General Specifications
Display Type: .......................................................................................................................5 digits VFD
A-to-D Conversion Rate:................................................................................... 45 conversions/seconds
Display Update:.........................5 user selections (100msec, 200msec, 300msec, 400msec, 500msec)
Overload:
20m
Ω
Range............................................................................................................ 99.95% of range
200m
Ω
thru 20k
Ω
.................................................................................................119.95% of range
Overload Indication: .........................................................................................flashes “
OVERLOAD
”
6
Terminal Configuration: ...............................................................................................Four-wire Kelvin
Test Current Polarity: ..............................................................................
..Positive
(flows
High
to
Low
)
Test Current Compliance Voltage:..................................................................................... 5V minimum
Settling Time:................................................................................................................ 300 milliseconds
2.4 Environmental and Power Requirements
Power Supply: ......................................... 115VAC or 230VAC ±10% @ 50Hz to 400Hz; 25VA max
Operating Temperature Range: ............................................................................................ 0
°
C to 50
°
C
Storage Temperature Range: .......................................................................................... -40
°
C to +85
°
C
2.5 Physical Specifications
Humidity:............................................................................... 80% RH max. @ 40
°
C (non-condensing)
Dimensions:.................................................................17"(43cm) W x 11½"(29.5cm) D x 4"(10cm) H
Weights:........................................................................ 10.36lbs (4.7kg) NET; 15lbs (7kg) SHIPPING
6
If the overload indication is on for approximately 10 seconds, the instrument will enter
Safe Mode
. See
section 7 of chapter 3 for details.