Rear Panel Functions
3-6
Force 2 Service Manual
Rear Panel Functions
On/Off Switch
– Press the toggle up to turn power on and down to shut
power off.
The Force 2 generator uses a universal type power entry module but is not
voltage selectable. Line fuse replacement is the following:
250V
8A
110-120V generators (-2 PCH)
250V
6.3A
220-240V generators (-8 PCH)
Bipolar Footswitch Receptacle
– This three-pin receptacle accepts a single-
pedal bipolar footswitch connector.
Monopolar Footswitch Receptacle
– This four-pin receptacle accepts a two-
pedal monopolar footswitch connector.
Equipotential Lug
– You can connect this lug to earth ground with an
equipotential grounding cable.
Audio Volume Control
– When the generator activates, it produces cut, coag,
and bipolar mode indicator tones. You adjust the volume of the tones
with this control. The volume of the audio alarm is not adjustable.
Summary of Contents for Force 2-2 PCH
Page 1: ...Service Manual Force 2 Electrosurgical Generator...
Page 10: ...Notes x Force 2 Service Manual...
Page 16: ...Notes 2 4 Force 2 Service Manual...
Page 32: ...Notes 4 10 Force 2 Service Manual...
Page 52: ...Typical Output Waveforms 6 12 Force 2 Service Manual Figure 6 5 Coag 300 Load...