Output Waveform
4-2
Force 2 Service Manual
Output Waveform
PCH Generator Output Characteristics
* An indication of a waveform’s ability to coagulate bleeders without a cutting effect
Output Configuration
Isolated output.
Cut
510 kHz sinusoid
Blend 1
510 kHz bursts of sinusoid at 50% duty cycle
recurring at 31 kHz
Blend 2
510 kHz bursts of sinusoid at 37.5% duty cycle
recurring at 31 kHz
Blend 3
510 kHz bursts of sinusoid at 25% duty cycle
recurring at 31 kHz
Coag
510 kHz damped sinusoidal bursts with a
repetition frequency of 31 kHz
Low Volt Coag
510 kHz bursts of sinusoid at 25% duty cycle
recurring at 31 kHz
Bipolar
510 kHz sinusoid, unmodulated
Maximum
(open circuit)
P-P voltage
Rated Load
(ohms)
Nominal
Power
at rated load
(watts)
Crest Factor*
at rated load
(typical)
Cut
3500
300
300
2.1 @ 100W
Blend 1
3800
300
250
3.4 @ 100W
Blend 2
4000
300
200
3.9 @ 100W
Blend 3
4000
300
150
4.7 @ 100W
Coag
7000
300
120
8.5 @ 50W
Low Voltage Coag
4000
300
99
4.7 @ 99W
Bipolar
1200
100
70
2.0 @ 40W