Calibrating the Force EZ-C Series Electrosurgical Generator
Setup, T
e
st
s, and Adjustment
s
Force EZ-C Series Service Manual
6-23
The following table summarizes the calibration steps and the values you can
adjust. Certain values are not adjustable, but require verification.
Calibration
Step
Description
Adjustable?
1
Force EZ-C Series Electrosurgical Generator
data
Generator model number
Master microcontroller software version
Feedback microcontroller software version
No (verify
value)
No (verify
value)
No (verify
value)
2
Calendar
Day
Month
Year
Yes
Yes
Yes
3
Clock
Hour
Minute
Yes
Yes
4
REM impedance
No (verify
values)
5
Current sense gain factor
Yes
6
Voltage sense gain factor
Yes
7
Reactance gain factor
Yes
8
ECON factor
No