Battery
Replacement Procedures
Force EZ-C Series Service Manual
8-3
Important
Battery
Equipment
•
Phillips screwdriver
Procedure
1.
Turn off the generator and disconnect the power cord from the wall receptacle.
2.
Loosen the four screws that secure the cover to the chassis. Lift the cover off
the chassis.
3.
Looking from the rear of the generator, locate the battery at the upper left
corner of the Control board.
4.
Grasp the battery and snap it out of the socket.
Internal memory loses all
stored data (refer to Section 4,
Internal Memory) when you
remove the battery .
5.
Install a new button-cell 3V battery. Refer to
Control Board Components
in
Section 10 for the Valleylab replacement battery or call the Valleylab Service
Center for information about approved battery replacements.
a.
Position the battery so that the positive side of the battery faces out
(visible).
b.
Snap the battery into the socket so it is firmly seated.
6.
Position the cover above the chassis and slide it down. Tighten the four screws
that secure the cover to the chassis.
Warning
Electric Shock Hazard
– To allow stored energy to dissipate after disconnecting
power, wait at least five minutes before replacing parts.
Caution
The generator contains electrostatic-sensitive components. When repairing the
generator, work at a static-control workstation. Wear a grounding strap when
handling electrostatic-sensitive components,
except
when working on an
energized generator. Handle circuit boards by their nonconductive edges. Use an
antistatic container for transport of electrostatic-sensitive components and circuit
boards.
Notice
Calibrate the generator after you install a new battery. Battery backed RAM loses
calibration values when you replace the battery. Refer to
Calibrating the Force
EZ-C Series Electrosurgical Generator
in Section 6 for instructions.