Standards and IEC Classifications
4-8
Force EZ-C Series Service Manual
Standards and IEC Classifications
The Force EZ-C Series Electrosurgical Generator meets all pertinent clauses of
the IEC 60601-1 second edition and IEC 60602-2-2 third edition.
Class I Equipment (IEC 60601-1)
Accessible conductive parts cannot become live in the event of a basic insulation
failure because of the way in which they are connected to the protective earth
conductor.
Type CF Equipment (IEC 60601-1)/Defibrillator Proof
ATTENTION
Consult accompanying documents.
The generator output is floating (isolated) with respect to
ground.
DANGER
Explosion risk if used with flammable anesthetics.
Rear Panel—To reduce the risk of electric shock, do not
remove the cover. Refer servicing to qualified service
personnel.
Front Panel—Dangerous Voltage Output.
Non-Ionizing Radiation
Classified with respect to electrical shock, fire, and
mechanical hazards only in accordance with UL60601-1 and
CAN/CSA C22.2 No. 601.1.
The Force EZ-C Series Electrosurgical Generator provides a
high degree of protection against electric shock, particularly
regarding allowable leakage currents. It is type CF isolated
(floating) output and may be used for procedures involving the
heart.
The Force EZ-C Series Electrosurgical Generator patient return
electrode terminal is protected from defibrillator discharge
according to ANSI/AAMI HF18 and IEC 60601-2-2.