Connections for Bipolar Surgery
6-4
Force EZ-C Series Service Manual
Connections for Bipolar Surgery
If you plan to use a footswitching bipolar instrument, you must connect a
footswitch. You may also use a footswitch to activate a handswitching
instrument.
Figure 6-1.
Bipolar connections (footswitch
activation from the Bipolar
Footswitch receptacle on the rear
panel)
Figure 6-2.
Bipolar connections (footswitch
activation from the Footswitch
receptacle on the front panel)
Warning
Electric Shock Hazard
–
• Do not connect wet accessories to the generator.
• Ensure that all accessories and adapters are correctly connected and that no
metal is exposed.
Caution
Accessories must be connected to the proper receptacle type. In particular,
bipolar accessories must be connected to the Bipolar Instrument receptacle only.
Improper connection of accessories may result in inadvertent generator activation
or a REM Contact Quality Monitor alarm.
footswitching
or
handswitching
bipolar footswitc
handswitching
or footswitching
instrument
Valleylab
monopolar
Press the Footswitch Selector button
until the left arrow indicator illuminates
green.